Milwaukee 2216-20 manual Required Equipment

Page 11

Required Equipment

Required equipment is listed in Table B. If the recommended models are not available, equipment with equivalent specifications may be used.

Repairs or servicing should be performed only by qualified personnel.

Table B. Required Equipment

Equipment

Required Characteristics

Recommended Model

 

 

 

Calibrator

AC Voltage Range: 0 750V AC

Fluke 5500 or Wavetek

 

Accuracy: ±0.07% (Basic)

9100 Calibrator or

 

Frequency Range: 40 1kHz

equipment

 

Accuracy: ±2%

 

 

DC Voltage Range: 0 1000V DC

 

 

Accuracy: ±0.006% (Basic)

 

 

Current Range: 0 20A

 

 

Accuracy: AC (40Hz to 1kHz): ±0.08% (Basic)

 

 

DC: ±0.02% (Basic)

 

 

Frequency Source: 0.5Hz 100kHz

 

 

Accuracy: ±0.001%

 

 

Amplitude: 0.5V p-p 600V p-p

 

 

Accuracy: ±5%

 

 

Resistance Range: 1Ω 100MΩ

 

 

Accuracy: ±0.03% (Basic)

 

 

Capacitance Range: 0.1uF 1100uF

 

 

Accuracy: ±0.10% (Basic)

 

 

 

 

9

Image 11
Contents MILWAUKEE ELECTRIC TOOL CORPORATION REPAIR SERVICECalibration INSTRUCTIONSPage TABLE OF CONTENTSIntroduction Precautions and Safety InformationThe Symbols Specific precautions SAFETYDescription SPECIFICATIONSGeneral specifications CharacteristicsMeasurement Characteristics 3 DC Current 6 Continuity 12Measuring Rate 3 times/sec 11Auto Power Off15Low Battery Voltage 2.3V ± 16Backlight 17Battery life ALKALINEDescription Physical and Environmental CharacteristicsCharacteristics Environmental characteristicsRequired Equipment Opening the Meter Case Basic MaintenanceReplacing the Battery Cleaning WarningTesting Fuses Fuse ReplacementPerformance Tests Testing the DisplayTesting the Voltage Function Testing the Resistance Function Testing the Ampere Function Level Table 7 Frequency TestStep SourceAOpen the bottom case Calibration ProcedureEnvironmental Condition Calibration ProcedureGCapacitance Calibration VR1 2.Tighten the screwEDCA Calibration VR7 FResistance 6kΩ Range Calibration VR6