Table 7 Frequency Test:

Table 7 Frequency Test:

Step

Source

Level

Waveform

1

8.0Vp-p

2.00Hz

Square Wave

 

 

 

 

2

8.0Vp-p

80.00Hz

Square Wave

 

 

 

 

3

8.0Vp-p

800.0Hz

Square Wave

 

 

 

 

4

8.0Vp-p

8.000kHz

Square Wave

 

 

 

 

5

10.0Vp-p

50.00kHz

Square Wave

 

 

 

 

6

230.0Vp-p

10.00Hz

Sine Wave

 

 

 

 

7

230.0Vp-p

100.0Hz

Sine Wave

 

 

 

 

Reading

1.98Hz to 2.02Hz

79.90Hz to 80.10Hz 799.0Hz to 801.0Hz 7.990kHz to 8.010kHz 49.93kHz to 50.07kHz 9.97Hz to 10.03Hz 49.93kHz to 50.07kHz

16

Image 18
Contents INSTRUCTIONS REPAIR SERVICECalibration MILWAUKEE ELECTRIC TOOL CORPORATIONTABLE OF CONTENTS PagePrecautions and Safety Information The SymbolsIntroduction SAFETY Specific precautionsCharacteristics SPECIFICATIONSGeneral specifications DescriptionMeasurement Characteristics 3 DC Current 6 Continuity 17Battery life ALKALINE 11Auto Power Off15Low Battery Voltage 2.3V ± 16Backlight 12Measuring Rate 3 times/secEnvironmental characteristics Physical and Environmental CharacteristicsCharacteristics DescriptionRequired Equipment Basic Maintenance Replacing the BatteryOpening the Meter Case Fuse Replacement WarningTesting Fuses CleaningTesting the Display Performance TestsTesting the Voltage Function Testing the Resistance Function Testing the Ampere Function Source Table 7 Frequency TestStep LevelCalibration Procedure Calibration ProcedureEnvironmental Condition AOpen the bottom caseFResistance 6kΩ Range Calibration VR6 2.Tighten the screwEDCA Calibration VR7 GCapacitance Calibration VR1