Milwaukee 2216-20 manual Testing the Voltage Function

Page 15

Testing the Voltage Function

To verify accuracy in the AC and DC voltage ranges, do the following:

1.Turn the rotary switch to “VTesting the Voltage Function” position.

2.Connect the Calibrator to the VΩ and COM inputs on the Meter.

3.Set the Calibrator for the voltage and frequency from step 1 to 12 in Table 1.

4.Compare the reading on the Meter display with the display reading shown in Table 1.

5.If the display reading falls outside of the range shown in Table 1, the Meter does not meet specification.

Table 1 AC Voltage Test:

Step

Input

Frequency

Reading

1

500mV

50Hz

494.7 to 505.3

 

 

 

 

2

500mV

500Hz

494.7 to 505.3

 

 

 

 

3

500mV

1kHz

489.7 to 510.3

 

 

 

 

4

5.000V

50Hz

4.947 to 5.053

 

 

 

 

5

5.000V

500Hz

4.947 to 5.053

 

 

 

 

6

5.000V

1kHz

4.897 to 5.103

 

 

 

 

7

50.00V

50Hz

49.47 to 50.53

 

 

 

 

8

50.00V

500Hz

49.47 to 50.53

 

 

 

 

9

50.00V

1kHz

48.97 to 51.03

 

 

 

 

10

500.0V

50Hz

494.7 to 505.3

 

 

 

 

11

500.0V

500Hz

494.7 to 505.3

 

 

 

 

12

500.0V

1kHz

489.7 to 510.3

 

 

 

 

6.Turn the rotary switch to “VManual background” position.

7.Set the calibration for the voltage from step 1 to 4 in Table 2.

8.Compare the reading on the Meter display with the display reading shown in Table 2.

9.If the display reading falls outside of the range shown in Table 2, the meter does not meet specification.

Table 2 DC Voltage Test:

Step

Input

Reading

1

500mV

497.3 to 502.7

 

 

 

2

5.000V

4.973 to 5.027

 

 

 

3

50.00V

49.73 to 50.27

 

 

 

4

500.0V

497.3 to 502.7

 

 

 

13

Image 15
Contents MILWAUKEE ELECTRIC TOOL CORPORATION REPAIR SERVICECalibration INSTRUCTIONSPage TABLE OF CONTENTSPrecautions and Safety Information The SymbolsIntroduction Specific precautions SAFETYDescription SPECIFICATIONSGeneral specifications CharacteristicsMeasurement Characteristics 3 DC Current 6 Continuity 12Measuring Rate 3 times/sec 11Auto Power Off15Low Battery Voltage 2.3V ± 16Backlight 17Battery life ALKALINEDescription Physical and Environmental CharacteristicsCharacteristics Environmental characteristicsRequired Equipment Basic Maintenance Replacing the BatteryOpening the Meter Case Cleaning WarningTesting Fuses Fuse ReplacementPerformance Tests Testing the DisplayTesting the Voltage Function Testing the Resistance Function Testing the Ampere Function Level Table 7 Frequency TestStep SourceAOpen the bottom case Calibration ProcedureEnvironmental Condition Calibration ProcedureGCapacitance Calibration VR1 2.Tighten the screwEDCA Calibration VR7 FResistance 6kΩ Range Calibration VR6