Milwaukee 2216-20 manual 11Auto Power Off, 12Measuring Rate 3 times/sec, 17Battery life ALKALINE

Page 9

(11)Auto Power Off

The meter automatically turns the power off after powering up and no operation for 20 ± 5 minutes. To power up, turn the knob back to OFF first.

(12)Measuring Rate: 3 times/sec.

(13)Fuse: 10A/600V Fast Action Fuse

(14)Battery: Alkaline 1.5V Battery * 2

(15)Low Battery Voltage: 2.3V ± 0.1V

(16)Backlight

The meter automatically turns the backlight off after lighting up and no operation for 10 ± 1 minutes.

(17)Battery life (ALKALINE):

Approx. 26 hours with backlight on. Approx. 360 hours with backlight off.

(18)Overvoltage Category IEC 61010-1 600V CAT. III

(19)Operating Temperature

-10°C ~ 30°C, <= 80%RH 30°C ~ 40°C, <= 75%RH 40°C ~ 50°C, <= 45%RH

(20)Storage Temperature

-40°C ~ +60°C, 0 ~ 80%RH (No batteries)

7

Image 9
Contents Calibration REPAIR SERVICEINSTRUCTIONS MILWAUKEE ELECTRIC TOOL CORPORATIONPage TABLE OF CONTENTSPrecautions and Safety Information The SymbolsIntroduction Specific precautions SAFETYGeneral specifications SPECIFICATIONSCharacteristics DescriptionMeasurement Characteristics 3 DC Current 6 Continuity 15Low Battery Voltage 2.3V ± 16Backlight 11Auto Power Off17Battery life ALKALINE 12Measuring Rate 3 times/secCharacteristics Physical and Environmental CharacteristicsEnvironmental characteristics DescriptionRequired Equipment Basic Maintenance Replacing the BatteryOpening the Meter Case Testing Fuses WarningFuse Replacement CleaningPerformance Tests Testing the DisplayTesting the Voltage Function Testing the Resistance Function Testing the Ampere Function Step Table 7 Frequency TestSource LevelEnvironmental Condition Calibration ProcedureCalibration Procedure AOpen the bottom caseEDCA Calibration VR7 2.Tighten the screwFResistance 6kΩ Range Calibration VR6 GCapacitance Calibration VR1