Bacharach 2772-0803 manual Cleaning Optics block

Page 38

REMOTE INTELLIGENT SENSOR - AREA MONITOR

Repeat several times, also with the gate open. If a can of pressurized air is available, carefully

position tube at the tape side of the gate and blow again several times. Suitable cans of air are usually available from photographic suppliers. A mechani- cal pump or bellows can be used but it may not be possible to blow from the tape side.

NOTE: Take care to ensure that any dirt, not trapped by the tape in the open gate, is removed. Clean both holes in twin hole blocks.

Carefully clean holes in block using a small amount of cotton wool twisted onto a sliver of wood. A normal sized cotton ‘bud’ or swab may be too large. If necessary, use a little residue free solvent spray on cotton wool as well. Afterwards, make sure that no debris has been left during cleaning process by blowing through again.

At this stage check the light levels to see if the cleaning has effected an adequate improvement. If it has, carefully reassemble the input tubes and, after replacing a serviceable cassette, recheck the light levels and put the system back into service.

CAUTION

Make sure that the input tubes are properly engaged in the block or the instrument will not monitor accurately (Figures 2-4 & 2-5).

4.10.5DIRECT CLEANING OF THE LED’S AND PHOTO DIODES. If the limited cleaning detailed above does not effect an adequate improve- ment in the light levels perform the following.

NOTE: This procedure requires removal of the optics block and should not be done without good reason.

Remove the cassette. If the input tubes have not already been removed, do so now. Take out the two slot-head screws securing the optics block to the mechanical chassis.

Carefully lift block without disconnecting the leads or putting them under undue strain. It is now possible to insert a small swab into the two holes in the block. This should be done from the tape side. Gently find the angled cavities at the end of

which are mounted the optic devices (Figure 4-4), clean both sides. Make sure that no lint or other particles are left behind, blow out if necessary.

Figure 4-4. Cleaning Optics block

4.10.6REPLACING THE OPTICS BLOCK. Care is required when replacing the Optics block so that it is correctly positioned relative to the other half. It is first necessary to position the tape ad- vance cam (on the left of the gate) so that it is vertical. This can be achieved in one of two ways:

With switch SW1 turned off, find the take up clutch on the mechanical chassis beneath the top plate as shown in Figure 4-5. Turn the clutch by moving the light colored gear at the bottom of the clutch assembly (this is furthest away from the top plate and it can be seen without taking the me- chanical chassis out of the RIS).

The clutch must be turned until stepping cam is vertical as shown in Figure 4-6. The gear is a little difficult to start turning because of the ratio.

The cam may also be positioned by turning switch SW1 ‘off ’, remove the connector at the top left of the unit as shown in Figure 4-5 and switch ‘on’ again, the stepping cam will turn continuously. By turning SW1 ‘off ’ at the appropriate point in the cam’s rotation, it may be positioned vertically as shown.

Figure 4-5. Finding Take Up Clutch

Page 4-4

Instruction 2772-0803

Image 38
Contents Instruction Remote Intelligent Sensor Area Monitor Declaration of Conformity Contents Contents Preparation Removal of the Mechanical Chassis Verifying That a GAS Alarm WAS Caused by GASCleaning the Sampling Input and Optics Block Replacing the Mechanical ChassisMonitored Concentration Levels High A1.1 Adjusting Light Levels on Phase 1 RIS UnitsRIS TEST/FAULT Parameter LOG Sheet Supplement aUser Selectable Alarm SET Points Continuous Diagnostics & System Test ModeBattery Backup Optional FeaturesPhysical Power Input RequirementsAccuracy OutputsKeyboard Internal BACK-UP BatteryDisplay EnvironmentalTable #1 RIS Catalog Numbers & Data System Description Basic PrinciplesDoor Open and Cover Removed from Back of Door Remote Intelligent Sensor Area Monitor Remote Intelligent Sensor Area Monitor Electrical Installation Battery ConnectionMechanical Installation UnpackingSystem Power Supply Consideration & Selection SpecificationSample Lines RIS TypeTools & Materials Required 1 Volt ConversionSystem Check Tape Cassette LoadingCheck and Adjust System For detailed instructions on adjustment Interruption of Test Mode Cycle. IfOption PCB Installation Without Terminal USESilkscreen Idents NAL Phase 2 Base Board Terminal Strips Remote Intelligent Sensor Area Monitor Remote Intelligent Sensor Area Monitor Remote Intelligent Sensor Area Monitor Complete Sampling Sequence Twin Track Tape SAMPLING. AtMinimum Sample Time CHANGE-OVER from Density to Time ModeDensity & Time Operating Modes Tape Cassette LifeFlashing System AlarmsConstant OpticsUsing the Keyboard DiagnosticsTest Mode Optional FeaturesPrinter Operation & USE Optional Feature Installing Test CardSwitch # 2 is not used and is left ‘ON SW1 TotalWhite RED Prdy Black Pprs BaudOFF Non-Bacharach Printer Interface Data Busy GroundAdjusting Light Levels Maintenance GeneralVerifying That a GAS Alarm WAS Caused by GAS Clearing a Spurious AlarmAdjusting the Door Switch Checking and Adjusting System Flow RateDisabling the ‘DOOR OPEN’ Alarm Cleaning the Sampling Input and Optics Block Measuring Pump CurrentCleaning Optics block Stepping Clamp Vertical Changing a PumpExterior Cleaning KEY Parameter ChecksMechanical Tightness Input Path CleaningTroubleshooting General Pump CheckExcessive Tape USE Light FaultsDoor Fault Tape BreakageRecommended Spares System Power SuppliesModel Dependent Parts Options and Supplies Service CentersCommon Parts Pennsylvania200 to HoldRemote Intelligent Sensor Area Monitor Instruction S-A3 S-A4 Instruction