Bacharach 2772-0803 manual Density & Time Operating Modes, CHANGE-OVER from Density to Time Mode

Page 28

REMOTE INTELLIGENT SENSOR - AREA MONITOR

This method eliminates the ‘zero drift’ that could otherwise occur with slight variation in the reflective value of the tape. It also ensures that system accuracy is not compromised when reusing the same tape spot. System accuracy is not compro- mised when reusing the same tape spot.

3.2DENSITY & TIME OPERATING MODES

Actual sampling takes place under control of one of two operating modes. The microprocessor automatically selects the appropriate mode in relation to the dynamic situation being monitored. The two modes are:

3.2.1DENSITY MODE. This operating mode is used when the sampled concentration is below a predetermined threshold. Under this condition, the gas sample period is a fixed four minutes.

During the four minutes, a sample is drawn into the system at a specified Flow Rate. The product of time and flow provides the measured volume used in the subsequent calculation of concentration.

The term ‘Density Mode’ is used because the sample volume is fixed and the variable used in the concentration calculation is stain density.

3.2.2TIME MODE. When the concentration level of the sampled gas exceeds the preset threshold value, the stain on the tape develops quicker. By measuring the time it takes to reach the threshold density, it is possible to calculate the sample concentration.

Using the ‘time mode’ provides real benefits; a reading and an attendant alarm are provided rapidly when concentrations are high, and the tape is never allowed to become saturated. Because the tape doesn’t saturate, the reaction is always in the linear region where the monitor’s accuracy is maximized.

An additional benefit is that the tape continues to act as a filter and prevents the sample getting past the optics block and contaminating the system beyond.

3.3CHANGE-OVER FROM DENSITY TO TIME MODE

With a 1,000 ppb instrument, the change-over from density to time mode occurs at 20 ppb. Other range instruments have change-over points in proportion. This dynamic control of sampling mode is automatic.

3.4MINIMUM SAMPLE TIME

The minimum sample time in the time mode is approximately 16 seconds. This reponse occurs when the monitored concentration has reached, or ex- ceeded, the maximum range of the instrument.

3.5TAPE CASSETTE LIFE

The cassette has a nominal life of 30 days. However the rate at which the system uses tape, and the consequent cassette life is dependenty mainly on the gas concentration monitored.

If the level is zero, or very low, a tape step will occur every 16 minutes (4 x four minutes). This is because the two tracks are each used for two consecutive samples of four minutes as explained in Section 3.1.2.

If the monitored concentration is higher, the tape steps every two sample periods; also the sample period itself can reduce below 4 minutes (dependent upon the concentration reached). A tape step will therefore occur at least every 8 minutes (2 x 4 minute sample periods) and more quickly if levels are high.

Infrequent high gas concentrations will not have much effect on cassette life, but if monitored levels are continuously significant, cassette life will be noticeably reduced.

In most applications, since monitoring is being done to confirm an expected zero, or very low gas concentration, cassette life will be close to the 30-day period specified.

3.6DATA POINT STORAGE

At the end of each sample interval the calcu- lated data point is stored, as are the calculated TWA’s. The 15 minute TWA , and the 8 hour TWA are both available for immediate display at a touch of the appropriate front panel key switch. The stored data points can be printed out as a record when the optional Printer option is fitted.

3.7SYSTEM DISPLAY

The displayed information, including symbols, or icons, depends upon the operating conditions. These and related information are as described below.

3.7.1SYSTEM NORMAL, GAS CONCENTRA- TION ZERO OR LOW. This is the usual operating condition and three pieces of information are dis- played.

Page 3-2

Instruction 2772-0803

Image 28
Contents Instruction Remote Intelligent Sensor Area Monitor Declaration of Conformity Contents Contents Verifying That a GAS Alarm WAS Caused by GAS Cleaning the Sampling Input and Optics BlockPreparation Removal of the Mechanical Chassis Replacing the Mechanical ChassisA1.1 Adjusting Light Levels on Phase 1 RIS Units RIS TEST/FAULT Parameter LOG SheetMonitored Concentration Levels High Supplement aContinuous Diagnostics & System Test Mode Battery BackupUser Selectable Alarm SET Points Optional FeaturesPower Input Requirements AccuracyPhysical OutputsInternal BACK-UP Battery DisplayKeyboard EnvironmentalTable #1 RIS Catalog Numbers & Data System Description Basic PrinciplesDoor Open and Cover Removed from Back of Door Remote Intelligent Sensor Area Monitor Remote Intelligent Sensor Area Monitor Battery Connection Mechanical InstallationElectrical Installation UnpackingSystem Power Supply Consideration & Selection SpecificationSample Lines RIS Type1 Volt Conversion System CheckTools & Materials Required Tape Cassette LoadingCheck and Adjust System For detailed instructions on adjustment Interruption of Test Mode Cycle. IfOption PCB Installation Silkscreen Idents NAL Terminal USEWithout Phase 2 Base Board Terminal Strips Remote Intelligent Sensor Area Monitor Remote Intelligent Sensor Area Monitor Remote Intelligent Sensor Area Monitor Complete Sampling Sequence Twin Track Tape SAMPLING. AtCHANGE-OVER from Density to Time Mode Density & Time Operating ModesMinimum Sample Time Tape Cassette LifeSystem Alarms ConstantFlashing OpticsDiagnostics Test ModeUsing the Keyboard Optional FeaturesPrinter Operation & USE Optional Feature Installing Test CardSwitch # 2 is not used and is left ‘ON SW1 TotalOFF BaudWhite RED Prdy Black Pprs Non-Bacharach Printer Interface Data Busy GroundMaintenance General Verifying That a GAS Alarm WAS Caused by GASAdjusting Light Levels Clearing a Spurious AlarmDisabling the ‘DOOR OPEN’ Alarm Checking and Adjusting System Flow RateAdjusting the Door Switch Cleaning the Sampling Input and Optics Block Measuring Pump CurrentCleaning Optics block Stepping Clamp Vertical Changing a PumpExterior Cleaning KEY Parameter ChecksInput Path Cleaning Troubleshooting GeneralMechanical Tightness Pump CheckLight Faults Door FaultExcessive Tape USE Tape BreakageModel Dependent Parts System Power SuppliesRecommended Spares Service Centers Common PartsOptions and Supplies Pennsylvania200 to HoldRemote Intelligent Sensor Area Monitor Instruction S-A3 S-A4 Instruction