Sierra Monitor Corporation 4501-06 Operation, Key ENTER, Model 4501-XXToxic Gas Sensor Module

Page 12
5. OPERATION

Model 4501-XX Toxic Gas Sensor Module

5. OPERATION

5.1INTRODUCTION

The 4501-XX Toxic Gas Sensor Module utilizes a visual menu system operated by means of a magnet. A magnet stick is supplied for this purpose. The menu system is used to calibrate the sensor module, and for maintenance and diagnostic procedures.

The following adjustments can be made to the sensor module:

Range adjustments (see section 5.4)

4mA and 20mA calibration adjustments (see section 5.4)

Span adjustment (see section 6.3.1)

5.2DATA ENTRY KEYPAD

The module menu system is operated by means of directing the magnet stick toward each of four independent

hall-effect magnetic switches. Each switch functions as if it is a manually activated panel key. The keys are located under the faceplate above and below the LCD and are labeled M , E , S and T as shown in Figure 5-1.

Key •Key  ENTER: MODE

Key Figure 5-1 ENTER

Key Manual background: UP (+)

Key Manual background: DOWN (-)

Figure 5-1

Face Plate with Operator Keypad

Page: 8

Image 12
Contents APPLICABILITY & EFFECTIVITY 4501-05FM Approved 4501-06FM Pending2-WireSeries MODELModel 4501-XXToxic Gas Sensor Module FM PERFORMANCE APPROVALModel 4501-05Hydrogen Sulfide Gas Sensor Module Sensor ModuleContents Model 4501-XXToxic Gas Sensor ModuleContents CABLING INSTALLATIONModel 4501-XXToxic Gas Sensor Module 1. PRODUCT DESCRIPTION PageModel 4501-XXToxic Gas Sensor Module 1.4.3SENSOR ASSEMBLY 1.5INTERCONNECT WIRING1.6POWER REQUIREMENTS Model 4501-XXToxic Gas Sensor ModuleModel 4501-XXToxic Gas Sensor Module Page2. CAUTIONS & WARNINGS Model 4501-XXToxic Gas Sensor ModulePage 3. QUICK START Model 4501-XXToxic Gas Sensor ModulePage 4. INSTALLATION Figure 4-20mA ConnectionModel 4501-XXToxic Gas Sensor Module PageFigure Connector Locations Page Model 4501-XXToxic Gas Sensor Module5. OPERATION Key ENTERFigure Face Plate with Operator Keypad Page Model 4501-XXToxic Gas Sensor ModuleFunction DisplayDescription ReferenceModel 4501-XXToxic Gas Sensor Module Sub-RoutineAFunction Display5.5DIAGNOSTICS SUB-MENU Diagnostics Sub MenuModel 4501-XXToxic Gas Sensor Module PageModel 4501-XXToxic Gas Sensor Module PageCALIBRATION PROCEDURE 6. CALIBRATIONModel Flow6.3 DIAGNOSTICS SUB-MENU Model 4501-XXToxic Gas Sensor ModulePage Key Function7. SERVICE 7.1SENSOR MODULE CONFIGURATIONENCLOSURE REPLACEMENT Figure Module Components PageTRANSMITTER REPLACEMENT SENSOR REPLACEMENTINSTALLATION INSPECTION Model 4501-XXToxic Gas Sensor ModuleWiring terminations clean and correct Model 4501-XXToxic Gas Sensor ModulePage Specifications APPENDIX A SPECIFICATIONS8. APPENDICES Model 4501-XXToxic Gas Sensor ModuleWarranty Limited warrantyModel 4501-XXToxic Gas Sensor Module PageAPPENDIX B MODEL NUMBERS & PARTS LIST Model 4501-XXToxic Gas Sensor ModulePage Sensor ModuleAPPENDIX C LIMITED WARRANTY Model 4501-XXToxic Gas Sensor ModulePage