Model
TRANSMITTER REPLACEMENT
The transmitter assembly should be replaced when it is determined that it is unreliable, noisy or cannot be adjusted for calibration. This may occur due to age, corrosion or failed components.
To replace the transmitter assembly:
a.Remove the cover of the main enclosure
b.Unscrew the two thumb screws in the top of the cover plate, lift the assembly and rotate 90o to relieve the wiring service loop
c.Unplug the sensor connector from the transmitter
d.Remove the two wires from J1 terminals (see figure
e.Reverse the preceding steps to install the new transmitter
f.Restore power and allow a minimum of 30 minutes for stabilization before
SENSOR REPLACEMENT
The gas sensor which is located inside the sensor assembly housing can be replaced without replacement of the housing. The gas sensor needs replacement when:
It is no longer possible to obtain correct Zero and Span values
The “LOW SENS” message appears after calibration
The sensor output signal is noisy, causing erroneous gas level readings
To replace the sensor:
a.Confirm that system power has been removed
b.Remove the gas sensor module enclosure cover,
c.Unscrew the two thumb screws in the top of the faceplate, lift the transmitter assembly and rotate 90o to relieve the wiring service loop.
d.Unplug the sensor connector from the transmitter
e.Unscrew the old sensor assembly from the enclosure conduit hub. Remove the sensor assembly with its harness
f.Unscrew sensor housing cover from the sensor
g.Carefully pull the old sensor straight up from the socket
h.Press the new sensor into the socket.
i.Reverse the preceding steps to install the sensor assembly.
j.Allow the new sensor to stabilize for a minimum of 30 minutes and then calibrate using the procedure in Section 6.
INSTALLATION INSPECTION
Prior to system
7.1.1CONTROLLER INSTALLATION
•Controller installed in conformance to manufacturer’s instruction manual recommendations.
7.1.2CABLING INSTALLATION
•All splices are soldered or via terminal block.
•Cabling is away from sources of electrical noise or RFI where possible.
7.1.3SENSOR MODULE INSTALLATION
•Module installation in conformance with this manual.
•Modules accessible for calibration.
Page: 16