Sierra Monitor Corporation 4501-XX, 4501-04 Transmitter Replacement, Sensor Replacement, Page

Page 20
TRANSMITTER REPLACEMENT

Model 4501-XX Toxic Gas Sensor Module

TRANSMITTER REPLACEMENT

The transmitter assembly should be replaced when it is determined that it is unreliable, noisy or cannot be adjusted for calibration. This may occur due to age, corrosion or failed components.

To replace the transmitter assembly:

a.Remove the cover of the main enclosure

b.Unscrew the two thumb screws in the top of the cover plate, lift the assembly and rotate 90o to relieve the wiring service loop

c.Unplug the sensor connector from the transmitter

d.Remove the two wires from J1 terminals (see figure 4-2)

e.Reverse the preceding steps to install the new transmitter

f.Restore power and allow a minimum of 30 minutes for stabilization before re-calibration

SENSOR REPLACEMENT

The gas sensor which is located inside the sensor assembly housing can be replaced without replacement of the housing. The gas sensor needs replacement when:

It is no longer possible to obtain correct Zero and Span values

The “LOW SENS” message appears after calibration

The sensor output signal is noisy, causing erroneous gas level readings

To replace the sensor:

a.Confirm that system power has been removed

b.Remove the gas sensor module enclosure cover,

c.Unscrew the two thumb screws in the top of the faceplate, lift the transmitter assembly and rotate 90o to relieve the wiring service loop.

d.Unplug the sensor connector from the transmitter

e.Unscrew the old sensor assembly from the enclosure conduit hub. Remove the sensor assembly with its harness

f.Unscrew sensor housing cover from the sensor

g.Carefully pull the old sensor straight up from the socket

h.Press the new sensor into the socket.

i.Reverse the preceding steps to install the sensor assembly.

j.Allow the new sensor to stabilize for a minimum of 30 minutes and then calibrate using the procedure in Section 6.

INSTALLATION INSPECTION

Prior to system start-up or trouble shooting, the entire system should be visually inspected. The following are guidelines for that inspection:

7.1.1CONTROLLER INSTALLATION

Controller installed in conformance to manufacturer’s instruction manual recommendations.

7.1.2CABLING INSTALLATION

All splices are soldered or via terminal block.

Cabling is away from sources of electrical noise or RFI where possible.

7.1.3SENSOR MODULE INSTALLATION

Module installation in conformance with this manual.

Modules accessible for calibration.

Page: 16

Image 20
Contents APPLICABILITY & EFFECTIVITY 4501-05FM Approved 4501-06FM Pending2-WireSeries MODELModel 4501-XXToxic Gas Sensor Module FM PERFORMANCE APPROVALModel 4501-05Hydrogen Sulfide Gas Sensor Module Sensor ModuleContents Model 4501-XXToxic Gas Sensor ModuleModel 4501-XXToxic Gas Sensor Module ContentsCABLING INSTALLATION Model 4501-XXToxic Gas Sensor Module 1. PRODUCT DESCRIPTIONPage 1.4.3SENSOR ASSEMBLY 1.5INTERCONNECT WIRING1.6POWER REQUIREMENTS Model 4501-XXToxic Gas Sensor ModuleModel 4501-XXToxic Gas Sensor Module PagePage 2. CAUTIONS & WARNINGSModel 4501-XXToxic Gas Sensor Module Page 3. QUICK STARTModel 4501-XXToxic Gas Sensor Module 4. INSTALLATION Figure 4-20mA ConnectionModel 4501-XXToxic Gas Sensor Module PageFigure Connector Locations Page Model 4501-XXToxic Gas Sensor Module5. OPERATION Key ENTERFigure Face Plate with Operator Keypad Page Model 4501-XXToxic Gas Sensor ModuleFunction DisplayDescription ReferenceModel 4501-XXToxic Gas Sensor Module Sub-RoutineAFunction Display5.5DIAGNOSTICS SUB-MENU Diagnostics Sub MenuModel 4501-XXToxic Gas Sensor Module PageModel 4501-XXToxic Gas Sensor Module PageCALIBRATION PROCEDURE 6. CALIBRATIONModel Flow6.3 DIAGNOSTICS SUB-MENU Model 4501-XXToxic Gas Sensor ModulePage Key Function7. SERVICE 7.1SENSOR MODULE CONFIGURATIONENCLOSURE REPLACEMENT Figure Module Components PageTRANSMITTER REPLACEMENT SENSOR REPLACEMENTINSTALLATION INSPECTION Model 4501-XXToxic Gas Sensor ModulePage Wiring terminations clean and correctModel 4501-XXToxic Gas Sensor Module Specifications APPENDIX A SPECIFICATIONS8. APPENDICES Model 4501-XXToxic Gas Sensor ModuleWarranty Limited warrantyModel 4501-XXToxic Gas Sensor Module PageAPPENDIX B MODEL NUMBERS & PARTS LIST Model 4501-XXToxic Gas Sensor ModulePage Sensor ModulePage APPENDIX C LIMITED WARRANTYModel 4501-XXToxic Gas Sensor Module