| Model | |
| 6.3 DIAGNOSTICS |
|
| ||||
|
|
|
|
|
|
|
|
| Key Function | Display | Description | Reference | |||
| M | E | Mode | H2S |
| Default Display (example | or |
| S | T | CALIB: |
| |||
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
| M | E | Enter | CAL 0 PPM |
| Banner: Apply zero gas, [E] when done |
|
| S | T |
|
| |||
|
|
|
|
|
| ||
|
|
|
|
|
| Operation: Confirm area clear of gas, or apply zero air to sensor. |
|
|
|
|
|
|
|
|
|
| M | E | Enter | Cal0PPM |
| There is a minimum time to calibrate at 0PPM. If [E] is hit prior to this time |
|
| S | T | Wait.XXX |
| has elapsed, a countdown screen will appear. |
| |
|
|
|
| ZERO |
| Zero gas setting completed (Message will appear on the screen and after 3 |
|
|
|
|
| DONE |
| seconds will go to the next menu - 25PPM Span) |
|
|
|
|
| XXPPM |
| Banner: [S] or [T] to change span value, [E] when done | |
|
|
|
|
|
|
|
|
|
|
|
| H2S |
| Operation: 148 second time out before sensor is returned to service. |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
| |
| M | E | Enter | H2S |
| Banner: Remove gas from sensor | |
| S | T |
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| M | E | Mode | CALIB |
| Operation: Hold magnet over [M] for 10 seconds to abort calibration |
|
| S | T | ABORTED |
|
| ||
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
| H2S |
| Default Display |
|
|
|
|
| XXX PPM |
|
| |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
| 25PPM SPAN | Banner: [S] or [T] to change span value, [E] when done |
| |
|
|
|
|
|
|
| |
| M | E | Down | C##PPM |
| Banner: Apply | Should be stable after 3 |
| S | T |
| minutes of gas exposure | |||
| M | E | Enter | H2S |
| Operation: Remove gas from sensor | If calibration failed, then |
| S | T | REMOVE GAS |
| |||
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
| H2S |
| Calibration successful |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| M | E | Enter | READ |
| Operation: Remove gas from sensor read manual, particularly section 6.3 and |
|
| S | T | MANUAL |
| 7.6, to check calibration procedures. |
| |
|
|
|
| H2S |
| Calibration Failed |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
| |
|
|
|
|
|
| Table |
|
|
|
|
|
|
| Calibration |
|
Page: 14