5.5DIAGNOSTICS SUB-MENU
Key Function | Display |
| Description | Reference | |||
| |||||||
| E | Mode | H2S |
| Default Display(example |
| |
S | T | XXX PPM |
| ||||
|
|
| |||||
M |
| Enter | H2S |
| Go to previous maintenance menu item |
| |
S | T | CALIB |
|
| |||
|
|
|
| ||||
M | E | Down | H2S |
|
|
| |
S |
| MAINT: |
|
|
| ||
|
|
|
|
| |||
M | E | Down | H2S |
|
|
| |
S |
| DIAG: |
|
|
| ||
|
|
|
|
| |||
M |
| Enter | DIAG: |
| Enter diagnostics |
| |
S | T | RawSig |
|
| |||
|
|
|
| ||||
M |
| Enter | Voltage reading of the output of the sensor element |
| |||
S | T | XXXXmV |
| ||||
|
|
| |||||
| E | Mode | DIAG: |
| Return to diagnostic menu |
| |
S | T | RawSig |
|
| |||
|
|
|
| ||||
M | E | Down | DIAG: |
|
|
| |
S |
| Ref |
|
|
| ||
|
|
|
|
| |||
M |
| Enter |
| Voltage reading of the output of the reference |
| ||
S | T | XXXXmV |
| ||||
|
|
| |||||
| E | Mode | DIAG: |
|
|
| |
S | T | Ref |
|
|
| ||
|
|
|
| ||||
M | E | Down | DIAG: |
|
|
| |
S |
| Temp |
|
|
| ||
|
|
|
|
| |||
M |
| Enter | TEMP |
| Temperature of sensor use [S] or [T] to switch between Celsius and Fahrenheit |
| |
S | T | XXXC |
|
| |||
|
|
|
| ||||
| E | Mode | DIAG: |
|
|
| |
S | T | Temp |
|
|
| ||
|
|
|
| ||||
M | E | Down | DIAG: |
|
|
| |
S |
|
|
|
| |||
|
|
|
|
| |||
M |
| Enter | Enter | ||||
S | T | Normal |
| ||||
|
|
|
| ||||
| E | Mode | DIAG: |
|
|
| |
S | T |
|
|
| |||
|
|
|
| ||||
M | E | Down | DIAG: |
|
|
| |
S |
| Version |
|
|
| ||
|
|
|
|
| |||
M |
| Enter |
| Version number of firmware |
| ||
S | T | VX.XX |
|
| |||
|
|
|
| ||||
| E | Mode | DIAG: |
|
|
| |
S | T | Version |
|
|
| ||
|
|
|
| ||||
M | E | Down | DIAG: |
|
|
| |
S |
| FactInit |
|
|
| ||
|
|
|
|
| |||
M |
| Enter | Accept? |
| Banner: Use [S] or [T] to change, [E] to accept | ||
S | T | N NO | Y | ||||
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
| |
|
|
| Normal |
|
|
M | E | Down | Press [E] to send 3.6 mA signal |
| |
S |
| 3.6 mA |
| ||
|
|
|
| ||
M | E | Down | Press [E] to send 4.0 mA signal |
| |
S |
| 4.0 mA |
| ||
|
|
|
| ||
M | E | Down | Press [E] to send 10.0 mA signal |
| |
S |
|
| |||
|
|
|
| ||
M | E | Down | Press [E] to send 20.0 mA signal |
| |
S |
| 20.0 mA |
| ||
|
|
|
|
M | E | Down |
| Accept? |
| If accepting [T] to change to yes and [E] to accept | |
S |
| N | YES | Y | |||
|
|
|
| ||||
M |
| Enter |
| Accept? |
|
|
|
S | T | N | YES | Y |
|
| |
|
|
| |||||
|
|
|
| FactInit |
| Factory initialization completed |
|
|
|
|
| Done |
|
| |
|
|
|
|
|
|
|
M |
| Enter | Accept? | If not accepting, [E] to accept NO | ||
S | T | N NO Y | ||||
|
|
| ||||
|
|
| FactInit | Factory initialization not done |
| |
|
|
| Aborted |
| ||
|
|
|
|
| ||
|
|
|
| Table |
| |
|
|
|
| Diagnostics Sub Menu |
|