Bacharach RDM800, HGM300 warranty Section, Clipping Fault Adjustment

Page 10

SECTION 5

Clipping Fault Adjustment

Locate R-34 Gain adjust potentiometer. Adjust DETECTOR voltage using the Diagnostic screen until the voltage is 4.24 volts. If the reference voltage is properly set, the clipping fault will go away after the unit samples the next zone.

Note - It may take several minutes for a voltage change to appear on the diagnostic screen due to the cycle time of the monitor from zone to zone.

If the clipping fault remains, the reference voltage must be verified and corrected if needed.

Locate TP1 (+) and TP2 (-) on the HGM300 main PC board (they are found to the left of R-33 the reference adjust potentiometer).

Connect a voltmeter to TP1 and TP2. Use R-33 (Ref. Adj.) to adjust the reference voltage to 5.000VDC +/- 0.002.

TP2 (-)

TP1 (+)

SECTION 6

R33 - Reference Adjust

5.000 VDC +/- 0.002 VDC

R34 - Gain

Adjust

Rev 3/2/09

10

Image 10
Contents HGM300/RDM800 Preventative Maintenance and Technical Support To Bypass the Warm UP Mode HGM300 / RDM800 Technical Support Notes HGM300 CPU ResetRDM800 CPU Reset Monitor will not Come OUT of Warm UPCommon RDM800 Indications Section Alarm LineSection No Flow Check Procedure Rev 3/2/09 No Flow Problems HOW to Force Taking NEW Pressure Readings Clipping Fault Adjustment Page Connector J9 HGM300 Voltage CheckAlarm Detail Screen HOW to Silence Alarms on the HGM300Sample Flow Ndir Source Rev 3/2/09 Materials Used Rev 3/2/09 Procedure Materials UsedRev 3/2/09 Rev 3/2/09 Contact Peter Pape