SECTION 11

TO ENTER A SYSTEM PASSWORD PROTECTED OR CHANGE A PASSWORD

1)The system must be up and communicating.

2)Press the button to the left of “HGM# 1”.

3)When the “PASSWORD” screen appears enter “313” as the password.

4)From the System Screen, go to the RDM SET UP Screen.

5)In the “PASSWORD” block, enter “000” or a new password. “000” will remove password protection completely.

Rev 3/2/09

20

Image 20
Contents HGM300/RDM800 Preventative Maintenance and Technical Support HGM300 / RDM800 Technical Support Notes HGM300 CPU Reset RDM800 CPU ResetTo Bypass the Warm UP Mode Monitor will not Come OUT of Warm UPCommon RDM800 Indications Section Alarm LineSection No Flow Check Procedure Rev 3/2/09 No Flow Problems HOW to Force Taking NEW Pressure Readings Clipping Fault Adjustment Page Connector J9 HGM300 Voltage CheckAlarm Detail Screen HOW to Silence Alarms on the HGM300Sample Flow Ndir Source Rev 3/2/09 Materials Used Rev 3/2/09 Procedure Materials UsedRev 3/2/09 Rev 3/2/09 Contact Peter Pape