Bacharach HGM300, RDM800 warranty Rev 3/2/09

Page 17

8.Verify that the unit is not in a purge cycle. This can be done by verifying a flow of air into zone 1. Read the (PRES) reading in the diagnostic screen (write down reading).

9.Connect the tubing from Reference Gas regulator to Zone 1 on the manifold of HGM 300. Note: The Fault light on HGM300 may illuminate until gas starts to flow from the reference bottle to unit. The pressure should be below the original reading (if reading does not drop the unit may be in the purge cycle. If so wait until purge cycle ends then continue). Adjust the variable regulator slowly clockwise until pressure is the same as original reading. Note: You may have to block or partially block the relief valve on the variable pressure regulator in order to achieve the appropriate pressure. Allow about 1 minute to settle then read the first PPM value in the diagnostics screen. The reading should be within +-10% of the concentration indicated on the reference bottle. NOTE: The unit may go into a purge cycle during the test process. If this happens, the PPM reading will drop to approximately 0PPM. This does not affect the verification results. However in order to obtain a valid PPM reading, the unit must be drawing from zone1.

10.Close reference by adjusting variable regulator counter clockwise until fully closed. Reset zone 1 information (distance, refrigerant, zone hold time) as desired for normal operation.

11.If reading in step 9 is not within the +-10% of the PPM reference bottle please check the following:

- Verify pressure from step 8

-Check that tubing is securely connected to the reference bottle and Zone 1 input port -Verify proper gas is selected (step 5)

-Verify unit is in hold for zone 1

-Verify verify proper pressure from variable gas regulator (step 9) -Verify that ample time was given for readings to stabilize

If the unit still does not measure within +-10% of the Reference Gas PPM, please contact the factory.

Rev 3/2/09

17

Image 17
Contents HGM300/RDM800 Preventative Maintenance and Technical Support RDM800 CPU Reset HGM300 / RDM800 Technical Support Notes HGM300 CPU ResetTo Bypass the Warm UP Mode Monitor will not Come OUT of Warm UPCommon RDM800 Indications Section Line AlarmSection No Flow Check Procedure Rev 3/2/09 No Flow Problems HOW to Force Taking NEW Pressure Readings Clipping Fault Adjustment Page HGM300 Voltage Check Connector J9HOW to Silence Alarms on the HGM300 Alarm Detail ScreenSample Flow Ndir Source Rev 3/2/09 Materials Used Rev 3/2/09 Materials Used ProcedureRev 3/2/09 Rev 3/2/09 Contact Peter Pape