Bacharach RDM800, HGM300 warranty No Flow Problems

Page 8

NO FLOW PROBLEMS

1)– Pinch off the tubing between the water trap and internal filter. Go to the diagnostic screen on the RDM800. Observe the “PRES”, “VAC” AND “AMB” readings at the bottom of the screen.

b)– If the pressure does not drop to normal vacuum levels (about 2 PSIA below pressure reading) the problem is in one of the following;

5)– Tubing from the filter to the bench.

6)- Tubing, “T” connector to pressure sensor.

7)– Tubing from “T” connector to I/R bench.

8)– Tubing from I/R bench to pump.

b)– If the pressure does drop to normal vacuum levels then go to the tubing between the solenoid assembly and the water trap and pinch it closed.

3)– If the pressure does not drop to the same level as the first reading the problem is in the water trap.

4)– If the pressure does drop to the same reading as the first reading the problem is most likely in either the tubing around the solenoid manifolds or the solenoids themselves.

Step “a” pinch tubing

Exhaust

NOTE –THERE ARE 2

ELBOWS ON THE

WATER TRAP. IF AN

ELBOW IS BROKEN,

THE MONITOR WILL

NOT RECOGNIZE MORE

THE 4 ZONES

Ensure that the drain valve is tight (Turn clockwise to tighten)

 

In line

 

 

“T”

 

Filter

 

 

connector

 

 

 

 

 

 

 

 

 

 

Pressure sensor

Solenoid manifold

Step “b” pinch tubing here

Pump

Water Trap

Check that “FLOAT” is free

I/R Bench

Rev 3/2/09

8

Image 8
Contents HGM300/RDM800 Preventative Maintenance and Technical Support HGM300 / RDM800 Technical Support Notes HGM300 CPU Reset RDM800 CPU ResetTo Bypass the Warm UP Mode Monitor will not Come OUT of Warm UPCommon RDM800 Indications Section Alarm LineSection No Flow Check Procedure Rev 3/2/09 No Flow Problems HOW to Force Taking NEW Pressure Readings Clipping Fault Adjustment Page Connector J9 HGM300 Voltage CheckAlarm Detail Screen HOW to Silence Alarms on the HGM300Sample Flow Ndir Source Rev 3/2/09 Materials Used Rev 3/2/09 Procedure Materials UsedRev 3/2/09 Rev 3/2/09 Contact Peter Pape