Bacharach HGM300, RDM800 warranty Rev 3/2/09

Page 7

ADDITIONAL NOTES ON “NO FLOW” PROBLEMS (THE FLOW RATE OF THE PUMP IS 2L/MIN)

A DIFFERENCE OF LESS THEN 0.1 PSIA BETWEEN AMBIENT AND PRESSURE WILL CAUSE A “N0 FLOW” FAULT.

A VACUUM OF 0.7 PSIA OR LESS BELOW PRESSURE WILL CAUSE A “N0 FLOW” FAULT.

TO OBSERVE PRESSURE READINGS ON THE DIAGNOSTIC SCREEN ZONE BY ZONE, YOU MUST PUT A ZONE INTO”ZONE HOLD” THEN GO TO THE DIAGNOSTIC SCREEN AND VERFY THAT YOU ARE OBSERVING THE PROPER ZONE AT THE TOP OF THE SCREEN.

IN LINE FILTER

PUMP

“T” CONNECTOR

PRESSURE SENSOR

I/R BENCH

SOLENOID

MANIFOLD

STEP”A”

PINCH TUBING

STEP “B” PINCH

TUBING HERE

EXHAUST

WATER TRAP

CHECK THAT “FLOAT” IS FREE

ENSURE THAT DRAIN PETCOCK IS TIGHT (TURN CLOCKWISE TO TIGHTEN

Rev 3/2/09

7

Image 7
Contents HGM300/RDM800 Preventative Maintenance and Technical Support Monitor will not Come OUT of Warm UP HGM300 / RDM800 Technical Support Notes HGM300 CPU ResetRDM800 CPU Reset To Bypass the Warm UP ModeCommon RDM800 Indications Section Line AlarmSection No Flow Check Procedure Rev 3/2/09 No Flow Problems HOW to Force Taking NEW Pressure Readings Clipping Fault Adjustment Page HGM300 Voltage Check Connector J9HOW to Silence Alarms on the HGM300 Alarm Detail ScreenSample Flow Ndir Source Rev 3/2/09 Materials Used Rev 3/2/09 Materials Used ProcedureRev 3/2/09 Rev 3/2/09 Contact Peter Pape