FREEZER COMPONENTS

1.Outer door: To open the outer door, grip the handle. On closing, lock the outer door latch completely.

2.Inner door: The operation of the inner door should be quick to minimize the temperature rise in chamber. Lock the inner door latch completely when the inner door is closed. The inner door is removable for cleaning or defrosting. See page 21 “Routine maintenance”.

3.Inner door latch: Always lock the inner door latch when the inner door is closed.

4.Outer door latch: Always lock the latch when the outer door is closed.

5.Magnetic door gasket: This provides a tight door seal and prevents cold air leak. Keep clean.

6.Air intake vent (grille): Do not block this vent to keep the proper cooling performance.

7.Caster: 4 casters are provided to facilitate moving of the cabinet. For the installation, adjust the leveling foot so that the front 2 casters cannot contact with the floor.

8.Leveling foot: These are screw bolts used to install and fix the unit. Adjust the height of the leveling feet by turning the screw bolts until 2 front casters are away from the floor.

9.Condenser filter (behind the grille): This filter prevents the condenser. The dusty filter may cause failure of refrigerating device. See page 20 “Routine maintenance” for the cleaning.

dust from accumulating on the Clean the filter once a month.

10.Space for temperature recorder: A temperature recorder (optional component) can be attached here. See page 27 “Temperature recorder”.

11.Lock: Turn clockwise to 180o with a key and the outer door is securely locked.

12.Control panel: Used for temperature setting and indication of operating status is displayed on the panel. See page 10 for details.

13.Access port (rear and bottom): This is used for leading a cable and sensor of a measuring equipment, or nozzle of back-up system to chamber.

14.Air intake port: After closing the outer door, if used to open soon. See page 17

15.Fixture (on back side): 2 fixtures serve as spacers between the cabinet and wall and also serve as hooks to fix the unit. See page 12 “Installation”.

16.Power switch: This is for turning ON/OFF the power to the unit. ON – “I” OFF – “$

17.Remote alarm terminal: This is used to notice an alarm condition of the unit to remote location. Refer to page 17 “Remote alarm terminal”.

18.Battery switch: This is a switch for a battery for power failure alarm. Normally, turn on this switch. Be sure to turn off this switch if the freezer is not in operating for the long period.

19.Fuse: 125 V, 3.15 A. Normal blow type.

-51 -

9

Image 10
Contents MDF-U7486S MDF-U7486SC MDF-U5486S MDF-U5486SC Contents Introduction Precautions for Safe Operation Unqualified personnel may cause electric shock or fire Abnormal operation may cause electric shock or fire Precautions for Safe Operation Environmental Conditions Freezer Components Freezer Components Control panel Installation Site Remove the packaging materials and tapes InstallationAdjust the leveling foot Fix the unitOperation after power failure START-UP of UnitChamber Temperature Setting KEY Lock FunctionAlarm Temperature Setting Setting of Alarm Resume Time Remote Alarm Terminal AIR Intake PortSetting of Compressor Delay Time Alarms & Safety Functions Alarms and safety functionsCleaning of cabinet Routine MaintenanceCleaning of condenser filter Defrosting of inside wall Cleaning of air intake port Condition Check / RemedyTrouble Shooting Malfunction Check/RemedyDisposal of Unit Recycle of batteryDecontamination of unit Location of a nickel-metal-hydride battery Disposal of BatteryDisposal of nickel-metal-hydride battery BACK-UP System Option Installation of MTR-480/MTR-L03 Temperature Recorder OptionInstallation of MTR-85H/MTR-G85 Interface Board OptionInventory Rack Option MDF-U5486SCSpecifications MDF-U5486SCMDF-U7486SC Performance Safety check sheet