Sanyo MDF-U5486SC instruction manual MDF-U7486SC

Page 31

SPECIFICATIONS

Name

Ultra-Low Temperature Freezer

Model

MDF-U7486S

 

MDF-U7486SC

External dimensions

W1130 x D875 x H1990 (mm)

Internal dimensions

W870 x D600 x H1280 (mm)

Effective capacity

 

668 L

 

Exterior

Painted steel

 

Interior

Painted steel

 

Outer door

Painted steel

 

Inner door

ABS resin panel with stainless frame, 2 doors

Shelf

Stainless steel, 3 shelves (adjustable)

 

W848 x D533 (mm), Load; 50 kg/shelf

Access port

17 mm diameter, 2 locations (back, bottom left corner)

Insulation

Rigid polyurethane foamed-in place

Compressor

Hermetic type, Output; 1100 W (high stage side), 1100 W (low stage side)

Evaporator

Tube on sheet type

Condenser

Fin and tube type (high stage side), Shell and tube type (low stage side)

Refrigerant

R-404A (high stage side), R-508B (low stage side)

Temperature controller

Microcomputer control system

Temperature display

Digital display

 

Thermal sensor

Platinum resistance (Pt 1000￿)

Alarm

High temp. alarm, Low temp. alarm, Power failure alarm

 

Filter check, Battery check, Fan motor check

Remote alarm contact

Allowable contact capacity:

30 V DC, 2 A

Battery

Nickel-metal-hydride battery, 6 V DC, 1100 mAh, Auto-recharge

Accessories

1 set of key, 1 scraper, 1 stick for air intake port cleaning

Weight

360 kg

 

365 kg

Voltage booster

None

 

Built-in

 

 

 

Optional component

Inventory rack (IR-220U)

 

Temperature recorder + Mounting kit (MTR-G85, MTR-85H+MDF-S3085)

 

Back-up system (CVK-UB2): LCO2

 

Interface board (MTR-480) , LAN interface board (MTR-L03)

 

Data acquisition system (MTR-5000)

Note:

 

 

 

 

• Design or specifications will be subject to change without notice.

*: Power source of the automatic temperature recorder shall be 220 V.

- 72 -

30

 

Image 31
Contents MDF-U7486S MDF-U7486SC MDF-U5486S MDF-U5486SC Contents Introduction Precautions for Safe Operation Unqualified personnel may cause electric shock or fire Abnormal operation may cause electric shock or fire Precautions for Safe Operation Environmental Conditions Freezer Components Freezer Components Control panel Installation Site Fix the unit InstallationAdjust the leveling foot Remove the packaging materials and tapesSTART-UP of Unit Operation after power failureKEY Lock Function Chamber Temperature SettingAlarm Temperature Setting Setting of Alarm Resume Time AIR Intake Port Remote Alarm TerminalSetting of Compressor Delay Time Alarms and safety functions Alarms & Safety FunctionsCleaning of cabinet Routine MaintenanceCleaning of condenser filter Defrosting of inside wall Condition Check / Remedy Cleaning of air intake portMalfunction Check/Remedy Trouble ShootingDisposal of Unit Recycle of batteryDecontamination of unit Location of a nickel-metal-hydride battery Disposal of BatteryDisposal of nickel-metal-hydride battery BACK-UP System Option Interface Board Option Temperature Recorder Option Installation of MTR-85H/MTR-G85 Installation of MTR-480/MTR-L03MDF-U5486SC Inventory Rack OptionMDF-U5486SC SpecificationsMDF-U7486SC Performance Safety check sheet