Sanyo MDF-U7486SC, MDF-U5486SC instruction manual Remote Alarm Terminal, AIR Intake Port

Page 18

REMOTE ALARM TERMINAL

The terminal of the remote alarm is installed at the lower left side of the unit. The alarm is outputted from this terminal. Contact capacity is DC 30V, 2 A.

Contact output:

 

between COM. and N.O.

between COM. and N.C.

At normal

Open

Close

At abnormal

Close

Open

Note:

The alarm buzzer is silenced by pressing alarm buzzer stop key (BUZZER) on the control panel during alarm condition. (A remote alarm is continuing the operation.)

The buzzer will be activated again after certain suspension if the alarm condition is continued.

The alarm is actuated when the power supply plug is disconnected from the outlet or the power switch is OFF.

REMOTE ALARM

N.C. COM N.O.

MAX DC30V 2A

AIR INTAKE PORT

When a door is closed and opened soon, a door does not sometimes open this product.

The warm air which went into the chamber is cooled down rapidly, and this is because air inside the chamber contracted.

When a door is closed and opened soon, a door is easy to open with the following process.

1.Turn the cap of the left side counterclockwise, and remove it.

2.Put the air into the chamber from about ten for twenty seconds, and open a door.

3.Close a cap if a door opens.

Cap

A door may not open in the above method when there are frost and ice in the air intake port. In that case, open a cap, and investigate the matter whether there is no frost inside the air intake port. Remove frost inside the air intake port in "the air intake port cleaning stick" of the accessories when there is frost. Clean a air intake port every month even when there is no frost inside the air intake port. Refer to page 22.

WARNING

For removing the frost of the air intake port, do not use a tool with sharp edge such as a knife or a screw driver.

- 59 -

17

Image 18
Contents MDF-U7486S MDF-U7486SC MDF-U5486S MDF-U5486SC Contents Introduction Precautions for Safe Operation Unqualified personnel may cause electric shock or fire Abnormal operation may cause electric shock or fire Precautions for Safe Operation Environmental Conditions Freezer Components Freezer Components Control panel Installation Site Remove the packaging materials and tapes InstallationAdjust the leveling foot Fix the unitOperation after power failure START-UP of UnitChamber Temperature Setting KEY Lock FunctionAlarm Temperature Setting Setting of Alarm Resume Time Remote Alarm Terminal AIR Intake PortSetting of Compressor Delay Time Alarms & Safety Functions Alarms and safety functionsRoutine Maintenance Cleaning of cabinetCleaning of condenser filter Defrosting of inside wall Cleaning of air intake port Condition Check / RemedyTrouble Shooting Malfunction Check/RemedyRecycle of battery Disposal of UnitDecontamination of unit Disposal of Battery Location of a nickel-metal-hydride batteryDisposal of nickel-metal-hydride battery BACK-UP System Option Installation of MTR-480/MTR-L03 Temperature Recorder OptionInstallation of MTR-85H/MTR-G85 Interface Board OptionInventory Rack Option MDF-U5486SCSpecifications MDF-U5486SCMDF-U7486SC Performance Safety check sheet

MDF-U5486SC, MDF-U5486S, MDF-U7486S, MDF-U7486SC specifications

Sanyo is a renowned brand in the field of refrigeration and climate control solutions, and its MDF series of laboratory freezers is a testament to its commitment to quality and innovation. The models MDF-U7486SC, MDF-U5486SC, MDF-U7486S, and MDF-U5486S are designed to provide optimal cold storage solutions for sensitive materials in laboratories, clinics, and research facilities.

The MDF-U7486SC and MDF-U5486SC models stand out with their superior temperature stability and precision. Both units feature a microprocessor control system that offers precise temperature regulation, ensuring that stored samples remain within the desired specifications. The cooling system utilizes advanced compressor technology, which enhances energy efficiency while maintaining a low noise operation, making them ideal for quiet laboratory environments.

With a storage capacity of 740 liters for the MDF-U7486SC and 540 liters for the MDF-U5486SC, these freezers provide ample space for large quantities of biological and pharmaceutical materials. The interiors are designed with adjustable shelving, allowing for flexible organization of samples and easy access.

In terms of security, these models come equipped with a comprehensive alarm system that alerts users to deviations in temperature and power failures. The built-in data logger feature records temperature changes, providing a crucial audit trail for compliance and quality assurance.

On the other hand, the MDF-U7486S and MDF-U5486S models offer similar benefits but cater to slightly different needs. The S versions emphasize energy efficiency coupled with a robust cooling performance. They are designed to minimize energy consumption while still delivering reliable cold storage solutions. With an easy-to-use control panel, users can quickly adjust settings and monitor performance, ensuring optimal operation.

All models are constructed with high-quality materials, including a durable exterior that can withstand the rigors of laboratory environments. The insulation is of premium quality, which contributes to maintaining stable internal temperatures while minimizing energy loss.

In conclusion, Sanyo's MDF-U7486SC, MDF-U5486SC, MDF-U7486S, and MDF-U5486S freezers are excellent choices for laboratories looking for reliable, efficient, and secure cold storage solutions. With advanced technologies, robust performance, and user-friendly features, these models meet the diverse needs of scientific research and medical storage while ensuring sample integrity and safety.