Sanyo MDF-U5486SC, MDF-U7486SC instruction manual Chamber Temperature Setting, KEY Lock Function

Page 15

CHAMBER TEMPERATURE SETTING

Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the desired temperature is -75oC.

Note: The unit is set at the factory that the chamber temperature -80oC.

Table 1. Basic operation sequence (Example: Chamber temperature -75oC)

 

Description of operation

Key operated

Indication after operation

 

 

 

The current chamber temperature is

 

 

 

 

 

 

 

 

1

Turn the power switch ON.

----

displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Press set key.

SET

The second digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When pressed, the figure of settable

3

Set to -75 with the numerical value

 

digit changes.

shift key and digit shift key.

 

When pressed, the settable digit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

is shifted.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set temperature is memorized and the

 

 

 

 

4

Press set key.

SET

current chamber temperature is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displayed.

Note:

The temperature set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation.

Although the value of the chamber temperature setting can range from -50oC to -90oC, the guaranteed temperature when there is no load is -85oC when the ambient temperature is 30oC.

The set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation. In this case, any setting before pressing set key (SET) is not memorized.

KEY LOCK FUNCTION

This unit is provided with the key lock function. When the key lock is ON, change of temperature setting through the key pad is not available. The key lock is set in OFF at the factory.

Display

Mode

Function

L

0

Key lock is OFF

Enable to change of temperature setting

L

1

Key lock is ON

Disable to change of temperature setting

Table 2. Procedure for key lock setting (change from key lock OFF to key lock ON)

 

Description of operation

Key operated

Indication after operation

1

 

----

The current chamber temperature is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displayed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Press digit shift key for 5 seconds.

 

The first digit is flashed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

Press numerical value shift key

 

When pressed, the figure of settable

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

and scroll the figure to 1.

 

digit changes.

 

 

 

 

 

 

 

 

 

 

 

The key lock is set to ON.

 

 

 

 

4

Press set key.

SET

The current chamber temperature is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

displayed.

The set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation. In this case, any setting before pressing set key (SET) is not memorized.

- 56 -

14

 

Image 15
Contents MDF-U7486S MDF-U7486SC MDF-U5486S MDF-U5486SC Contents Introduction Precautions for Safe Operation Unqualified personnel may cause electric shock or fire Abnormal operation may cause electric shock or fire Precautions for Safe Operation Environmental Conditions Freezer Components Freezer Components Control panel Installation Site Fix the unit InstallationAdjust the leveling foot Remove the packaging materials and tapesSTART-UP of Unit Operation after power failureKEY Lock Function Chamber Temperature SettingAlarm Temperature Setting Setting of Alarm Resume Time AIR Intake Port Remote Alarm TerminalSetting of Compressor Delay Time Alarms and safety functions Alarms & Safety FunctionsRoutine Maintenance Cleaning of cabinetCleaning of condenser filter Defrosting of inside wall Condition Check / Remedy Cleaning of air intake portMalfunction Check/Remedy Trouble ShootingRecycle of battery Disposal of UnitDecontamination of unit Disposal of Battery Location of a nickel-metal-hydride batteryDisposal of nickel-metal-hydride battery BACK-UP System Option Interface Board Option Temperature Recorder OptionInstallation of MTR-85H/MTR-G85 Installation of MTR-480/MTR-L03MDF-U5486SC Inventory Rack OptionMDF-U5486SC SpecificationsMDF-U7486SC Performance Safety check sheet

MDF-U5486SC, MDF-U5486S, MDF-U7486S, MDF-U7486SC specifications

Sanyo is a renowned brand in the field of refrigeration and climate control solutions, and its MDF series of laboratory freezers is a testament to its commitment to quality and innovation. The models MDF-U7486SC, MDF-U5486SC, MDF-U7486S, and MDF-U5486S are designed to provide optimal cold storage solutions for sensitive materials in laboratories, clinics, and research facilities.

The MDF-U7486SC and MDF-U5486SC models stand out with their superior temperature stability and precision. Both units feature a microprocessor control system that offers precise temperature regulation, ensuring that stored samples remain within the desired specifications. The cooling system utilizes advanced compressor technology, which enhances energy efficiency while maintaining a low noise operation, making them ideal for quiet laboratory environments.

With a storage capacity of 740 liters for the MDF-U7486SC and 540 liters for the MDF-U5486SC, these freezers provide ample space for large quantities of biological and pharmaceutical materials. The interiors are designed with adjustable shelving, allowing for flexible organization of samples and easy access.

In terms of security, these models come equipped with a comprehensive alarm system that alerts users to deviations in temperature and power failures. The built-in data logger feature records temperature changes, providing a crucial audit trail for compliance and quality assurance.

On the other hand, the MDF-U7486S and MDF-U5486S models offer similar benefits but cater to slightly different needs. The S versions emphasize energy efficiency coupled with a robust cooling performance. They are designed to minimize energy consumption while still delivering reliable cold storage solutions. With an easy-to-use control panel, users can quickly adjust settings and monitor performance, ensuring optimal operation.

All models are constructed with high-quality materials, including a durable exterior that can withstand the rigors of laboratory environments. The insulation is of premium quality, which contributes to maintaining stable internal temperatures while minimizing energy loss.

In conclusion, Sanyo's MDF-U7486SC, MDF-U5486SC, MDF-U7486S, and MDF-U5486S freezers are excellent choices for laboratories looking for reliable, efficient, and secure cold storage solutions. With advanced technologies, robust performance, and user-friendly features, these models meet the diverse needs of scientific research and medical storage while ensuring sample integrity and safety.