Fluke 943-227-15, 943-27 manual Set Point Error Codes, Software Calculations

Page 24

Operation

Configuration 3

3.5 Set Point Error Codes

Failure of the 943-227-17 preamplifier may result in the following error codes being displayed on the controlling device:

Code E0009 is normally used to identify a Loop communication failure. The error code may be due to a baud rate mismatch between the preamplifier and the UDR or 960 system communication board, a failure on either the preamplifier or UDR communication board. To clear the error code, the source of the error must be corrected.

3.6 Software Calculations

The 943-227-17 continuously converts the current output of the detector to a digital value, controls the gain of the internal amplifier, performs periodic self checks, communicates with the UDR or Model 960 controller and provides a radiation value to the UDR of Model 960. An asynchronous, serial communication interface, processed through optically isolated drivers/receivers is utilized. The digital value is transmitted to the associated UDR or 960 System controller for further processing. Refer to the associated controller manual for further information. .

3-6

Image 24 Contents
Victoreen Fluke Biomedical Table of Contents Blank General Information General DescriptionApplication Specifications Preamplifier Variations Application Detector ReadoutFeature Specification Preamplifier Power Switch Normal Switch Function OFF Selects 300 baud 1/ON Selects 4800 baud normal settingJumper Normal Setting Function Alternate Setting Jumper Normal Setting Function Cal. PositionTB1 943-227-15 Description Connects To 977-210-10MJ4-1 Detector ConnectionsAuxiliary Equipment Recommended Spare PartsReceiving Inspection Part Number Description Used OnStorage Blank Theory of Operation Functional DescriptionFunction Hex Address Theory of Operation Main Circuit Boards Bit Position DescriptionTest Point Function RS232 TransmitVictoreen Communications Loop 943-227-15VL RS232C CommunicationsJumper Normal Function Alternate Setting Enables Victoreen LoopJumper Normal Function Cal Setting Position Installation OperationPreamplifier Mounting Electrical Interface Set-up Digital Preamplifier Fail Alarm Power FailureOperation Normal OperationConfiguration Functions To B Enables Victoreen Loop To C Enables RS232C NormalTo B Normal Operation To C PGA Offset Adjust Firmware VersionSet Point Error Codes Software CalculationsBlank Maintenance, Calibration and Troubleshooting MaintenanceElectronic Adjustments Electronic CalibrationTroubleshooting Troubleshooting / AdjustmentsConnector Designations Connector DESIGNATIONS, 943-227-15 and 943-227-15VLSwitch and Jumper POSITIONS, 943-227-15 and 943-227-15VL Applicable Drawings 943-227-15VL Related Manuals Not Supplied with this documentBlank Bill of Materials Bill of Materials- 943-227-15 and 943-227-15VLPart Number Description Blank Cable Termination Instructions Cable Termination InstructionsDetector Cable Termination Digital Preamplifier HV Connector Notes HV Digital Preamplifier Cable Termination Instructions D Digital Preamplifier Appendix Cable Termination Instructions Digital Preamplifier Modification Sheets, Engineering Instructions Modification Sheets, Engineering InstructionsBlank Blank Fluke Biomedical