Fluke 943-27, 943-227-15 manual General Information, General Description, Application

Page 5

General Information

General Description 1

Section 1

General Information

1.1 General Description

The Victoreen® Model 943-227-15 Digital Preamplifier, is designed to operate with the Model 943-27, Current Mode Beta Scintillation detector and the Model 942A-200C Series Digital Ratemeter (UDR) or the Model 960 Series Digital Process Control system. When connected to the beta sensitive scintillation detector, it comprises a monitoring system which operates in the range of 10-3to 105 uCi/cc (or 1E0 to 1E8

Pico amperes). The UDR provides display, control, and annunciator functions for the monitoring system. The preamplifier contains a microprocessor controlled, auto-zeroing, integrating electrometer, a programmable gain amplifier, an analog-to-digital converter and an asynchronous serial communications interface. Communications with the UDR is accomplished via the optically isolated Victoreen Serial Communication Loop driver/receiver circuitry.

Interconnection of the detector and preamplifier is accomplished utilizing two, six (6) foot coaxial cables. One cable supplies the high voltage for the detector (negative polarity). The other cable is used for the detector output signal. Interconnection between the preamplifier and the UDR or 960 system is accomplished via a six-conductor cable and a high voltage coaxial cable.

The detector operates at a potential of approximately (-) 500 VDC, and produces a nominal DC output current of 1 x 10-12amperes to 1 x 10-4amperes. Depending on sampling geometry and detector set up, this corresponds to a range of 1 x 10-3to 1 x 10+5 uCi/cc.

The preamplifier is a self-contained electronics unit that processes the DC input signal from the detector, digitizes the signal, and transmits the digitized value to the readout via a modified RS232C compatible serial data communication port. It contains two printed circuit boards (mounted on the rear of the enclosure cover, a terminal block for terminating the field cable from its associated controller and two connectors for terminating the detector Signal and HV coaxial cables. The larger of the two printed circuit boards is the 977-210-10M Serial Interface board. This board controls the communications with the controller. The smaller board is the 977-200-15M Electrometer. It contains the microprocessor and associated circuitry for converting the detector output current into a digital value.

For additional information on the detector or readout device, refer to the instruction manual for the Model 943-27-31 detector, the Model 942A-200C UDR, or the applicable Model 960 System manual.

1.2 Application

The Model 943-227-15 Digital Preamplifier is used with the Victoreen 942A-200C or Model 960 microprocessor based radiation monitors. The Model 942A-200C or 960 system provides the +15 VDC preamplifier electronics power, (-) 500 VDC detector power and the RS232C serial communications support. The digital preamplifier is available in two serial port configurations. For applications less than 50 feet, the Model 943-227-15, configured for RS232C communications, is used. For applications requiring signal transmission distances of up to 1 mile, the Model 943-227-15VL, configured for Victoreen Loop communications is used. Table 1-1 is a list of available configurations and compatible 9XX series equipment. Please contact Fluke Biomedical Radiation Management Services for additional information.

1-1

Image 5 Contents
Victoreen Fluke Biomedical Table of Contents Blank Application General InformationGeneral Description Preamplifier Variations Application Detector Readout SpecificationsFeature Specification Preamplifier Power OFF Selects 300 baud 1/ON Selects 4800 baud normal setting Switch Normal Switch FunctionJumper Normal Setting Function Alternate Setting Jumper Normal Setting Function Cal. Position943-227-15 Description Connects To 977-210-10M TB1J4-1 Detector ConnectionsRecommended Spare Parts Auxiliary EquipmentReceiving Inspection Part Number Description Used OnStorage Blank Functional Description Theory of OperationFunction Hex Address Bit Position Description Theory of Operation Main Circuit BoardsRS232 Transmit Test Point FunctionRS232C Communications Victoreen Communications Loop 943-227-15VLJumper Normal Function Cal Setting Position Jumper Normal Function Alternate SettingEnables Victoreen Loop Operation InstallationPreamplifier Mounting Electrical InterfaceSet-up Digital Preamplifier Power Failure Fail AlarmOperation Normal OperationTo B Enables Victoreen Loop To C Enables RS232C Normal Configuration FunctionsTo B Normal Operation To C PGA Offset Adjust Firmware VersionSoftware Calculations Set Point Error CodesBlank Maintenance Maintenance, Calibration and TroubleshootingElectronic Adjustments Electronic CalibrationTroubleshooting / Adjustments TroubleshootingConnector DESIGNATIONS, 943-227-15 and 943-227-15VL Connector DesignationsSwitch and Jumper POSITIONS, 943-227-15 and 943-227-15VL Related Manuals Not Supplied with this document Applicable Drawings 943-227-15VLBlank Part Number Description Bill of MaterialsBill of Materials- 943-227-15 and 943-227-15VL Blank Detector Cable Termination Cable Termination InstructionsCable Termination Instructions Digital Preamplifier HV Connector Notes HV Digital Preamplifier Cable Termination Instructions D Digital Preamplifier Appendix Cable Termination Instructions Digital Preamplifier Modification Sheets, Engineering Instructions Modification Sheets, Engineering InstructionsBlank Blank Fluke Biomedical