Fluke 943-27 manual Victoreen Communications Loop 943-227-15VL, RS232C Communications

Page 17

Theory of Operation

Communications 2

Victoreen Communications Loop – 943-227-15VL

Communication between the detector preamplifier and the UDR or 960 controller may be performed via Victoreen loop, a serial asynchronous communications interface. The communication utilizes Victoreen protocol and is intended for use in electrically hostile environments where high noise immunity is required. Noise immunity is achieved using differential line driving, and electrically isolating the end of each transmit/receive pair with opto-isolators. Using a ±15 Vdc communications power supply, each 0 to 1 transition is equivalent to a 30-volt swing in loop voltage. Refer to the Model 942-200-80-1 Communications Board instruction manual for further information.

Jumper JP1 controls the communication communications method employed. JP1, A-B selects Victoreen loop communications.

RS232C Communications – 943-227-15

Communication between the detector preamplifier and the UDR or 960 controller may be performed via RS232C serial asynchronous communications interface. Jumper JP1 controls the communication communications method employed.

JP1, B-C selects RS232C communications.

2.5 Preamplifier Switch & Jumper Settings

Switch and jumper settings for the preamplifier are summarized in Table 2-7, 2-8, and 2-9.

Table 2-7. Preamplifier Serial Interface Board 977-210-10M Switch Settings

Switch

Normal

Switch Function

 

Op.

 

SW1

Open

Microprocessor Reset switch, momentary contact

 

 

 

SW2

N/A

Not supplied

 

 

 

SW3-1

ON

OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)

SW3-2

OFF

OFF • (No function is implemented in the 977-210-10M)

SW3-3

OFF

OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for

 

 

factory PGA and Auto Zero set-up only)

SW3-4

OFF

OFF • (No function is implemented in the 977-210-10M)

NOTE

SW3 is a 4-position DIP switch on the 977-210-10M serial interface board.

2-5

Image 17
Contents Victoreen Fluke Biomedical Table of Contents Blank Application General InformationGeneral Description Preamplifier Variations Application Detector Readout SpecificationsFeature Specification Preamplifier Power OFF Selects 300 baud 1/ON Selects 4800 baud normal setting Switch Normal Switch FunctionJumper Normal Setting Function Alternate Setting Jumper Normal Setting Function Cal. Position943-227-15 Description Connects To 977-210-10M TB1J4-1 Detector ConnectionsRecommended Spare Parts Auxiliary EquipmentReceiving Inspection Part Number Description Used OnStorage Blank Functional Description Theory of OperationFunction Hex Address Bit Position Description Theory of Operation Main Circuit BoardsRS232 Transmit Test Point FunctionRS232C Communications Victoreen Communications Loop 943-227-15VLJumper Normal Function Cal Setting Position Jumper Normal Function Alternate SettingEnables Victoreen Loop Operation InstallationPreamplifier Mounting Electrical InterfaceSet-up Digital Preamplifier Power Failure Fail AlarmOperation Normal OperationTo B Enables Victoreen Loop To C Enables RS232C Normal Configuration FunctionsTo B Normal Operation To C PGA Offset Adjust Firmware VersionSoftware Calculations Set Point Error CodesBlank Maintenance Maintenance, Calibration and TroubleshootingElectronic Adjustments Electronic CalibrationTroubleshooting / Adjustments TroubleshootingConnector DESIGNATIONS, 943-227-15 and 943-227-15VL Connector DesignationsSwitch and Jumper POSITIONS, 943-227-15 and 943-227-15VL Related Manuals Not Supplied with this document Applicable Drawings 943-227-15VLBlank Part Number Description Bill of MaterialsBill of Materials- 943-227-15 and 943-227-15VL Blank Detector Cable Termination Cable Termination InstructionsCable Termination Instructions Digital Preamplifier HV Connector Notes HV Digital Preamplifier Cable Termination Instructions D Digital Preamplifier Appendix Cable Termination Instructions Digital Preamplifier Modification Sheets, Engineering Instructions Modification Sheets, Engineering InstructionsBlank Blank Fluke Biomedical

943-227-15, 943-27 specifications

The Fluke 943-27 and Fluke 943-227-15 are advanced electrical testing tools designed to meet the demands of professionals in industrial and commercial settings. Recognized for their accuracy and reliability, these devices come loaded with essential features tailored to ensure optimal performance during electrical measurements.

The Fluke 943-27 is a compact yet powerful multimeter that excels in providing precise voltage and current readings. It is equipped with a large, backlit display that ensures visibility in low-light environments. The device supports a wide range of testing parameters, including AC and DC voltage, resistance, continuity, and capacitance measurements, making it invaluable for troubleshooting electrical circuits.

On the other hand, the Fluke 943-227-15 offers additional capabilities, including advanced data logging features. This multimeter allows users to record and analyze measurements over time, which is particularly useful for identifying intermittent issues in electrical systems. Its connectivity options enable seamless data transfer to a computer or mobile device, facilitating easy documentation and reporting.

Both models illustrate a robust construction built to endure tough working conditions. They meet stringent safety standards, which is critical for protecting users from electrical hazards. The ergonomic design ensures comfort during prolonged use, and features like overmolding enhance grip and durability.

One of the standout technologies in these multimeters is Flukeā€™s True-RMS measurement capability. This feature ensures accurate readings of non-linear loads, which are common in modern electrical systems with variable frequency drives and other electronic devices. Additionally, the built-in low-pass filter technology reduces noise in measurements, ensuring precision even in environments with significant electrical interference.

The Fluke 943-27 and 943-227-15 are also designed to be user-friendly. They feature intuitive controls and a straightforward user interface, allowing technicians to navigate functions effortlessly. The ability to perform automatic ranging simplifies measurements, eliminating the potential for errors when setting the range manually.

In conclusion, the Fluke 943-27 and 943-227-15 represent essential tools for electrical technicians who require versatility, accuracy, and durability in their testing equipment. With advanced features, reliable technologies, and a user-centric design, these multimeters not only meet but exceed the expectations of professionals in the field.