Maintenance, Calibration and Troubleshooting

Maintenance 4

Section 4

Maintenance, Calibration and Troubleshooting

4.1 Maintenance

The 943-227-15 Digital Preamplifier is designed to operate for extended periods of time with no scheduled maintenance required. Operation may be verified by periodically actuating the check source (provided with the detector and sampling assembly) and observing the response of the unit. If the response varies by more than 50% of the normal value, further troubleshooting may be required. If a problem develops, troubleshoot the unit per Section 4.3 and the drawings in Appendix B.

Whenever the preamplifier enclosure is opened, the condition of the internal desiccant package should be checked. The desiccant bag is provided in the enclosure to absorb moisture. The desiccant bags should be checked for moisture absorption prior to placing into service, or whenever the electronics enclosure cover is opened. Each bag contains a color indicator stripe that turns red when it should be replaced.

4.2 Electronic Calibration

Electronic Adjustments

There are no user adjustments available on the 943-227-15 Digital Preamplifier. The Programmable Gain Amplifier and Auto Zero are factory adjustments, requiring the use of a factory test PROM and test fixture. We strongly recommend the unit be returned to the factory for repair or adjustment.

Table 4-2 lists the electronic adjustments for the 943-227-15 digital preamplifier. The procedures for making these adjustments are contained in the factory Calibration Procedure listed below. Note that special test firmware is required to perform the electrical adjustments.

Table 4-1. Electronic Adjustments

Potentiometer

Adjustment (943-227-15)

Tolerance

R11

Programmable Gain Amplifier Offset Adjustment

± 2 mV

R12

Auto Zero Set-Up Adjustment

± 10 mV

NOTE

Refer to the drawings in Appendix B to locate adjustment and/or test points, if required.

The follow factory calibration and test procedures are listed for reference only.

Document

Description

TP943-227-15VL

Test Procedure, Victoreen Loop Communications

TP943-227-15

Test Procedure, RS232 communications

TP977-200-15M

Electrometer PCB Test Procedure

4-1

Image 26 Contents
Victoreen Fluke Biomedical Table of Contents Blank Application General InformationGeneral Description Specifications Preamplifier Variations Application Detector ReadoutFeature Specification Preamplifier Power Jumper Normal Setting Function Alternate Setting Switch Normal Switch FunctionOFF Selects 300 baud 1/ON Selects 4800 baud normal setting Jumper Normal Setting Function Cal. PositionJ4-1 TB1943-227-15 Description Connects To 977-210-10M Detector ConnectionsReceiving Inspection Auxiliary EquipmentRecommended Spare Parts Part Number Description Used OnStorage Blank Theory of Operation Functional DescriptionFunction Hex Address Theory of Operation Main Circuit Boards Bit Position DescriptionTest Point Function RS232 TransmitVictoreen Communications Loop 943-227-15VL RS232C CommunicationsJumper Normal Function Cal Setting Position Jumper Normal Function Alternate SettingEnables Victoreen Loop Preamplifier Mounting InstallationOperation Electrical InterfaceSet-up Digital Preamplifier Operation Fail Alarm Power Failure Normal OperationTo B Normal Operation To C PGA Offset Adjust Configuration FunctionsTo B Enables Victoreen Loop To C Enables RS232C Normal Firmware VersionSet Point Error Codes Software CalculationsBlank Electronic Adjustments Maintenance, Calibration and TroubleshootingMaintenance Electronic CalibrationTroubleshooting Troubleshooting / AdjustmentsConnector Designations Connector DESIGNATIONS, 943-227-15 and 943-227-15VLSwitch and Jumper POSITIONS, 943-227-15 and 943-227-15VL Applicable Drawings 943-227-15VL Related Manuals Not Supplied with this documentBlank Part Number Description Bill of MaterialsBill of Materials- 943-227-15 and 943-227-15VL Blank Detector Cable Termination Cable Termination InstructionsCable Termination Instructions Digital Preamplifier HV Connector Notes HV Digital Preamplifier Cable Termination Instructions D Digital Preamplifier Appendix Cable Termination Instructions Digital Preamplifier Modification Sheets, Engineering Instructions Modification Sheets, Engineering InstructionsBlank Blank Fluke Biomedical