GE CTR 80 manual Bath Tank and Lid, Back Panel

Page 28

7 Parts and Controls

4)The heater mode is a red light emitting diode (LED). This indicator lets the user visually see the ratio of heating to cooling. When the indicator is lit the heater is on, and when it is off the heater is off and the bath is cooling.

5)The cooling switch turns on the refrigeration for control below 50°C and rapid cool down.

7.2Bath Tank and Lid

The bath tank and lid assembly includes: the tank, the control probe, the stir- ring motor, the access hole, and the access hole cover.

The bath tank is constructed of stainless steel. It is very resistant to oxida- tion in the presence of most chemicals and over a wide range of tempera- tures.

The control probe provides the temperature feedback signal to the con- troller allowing the controller to maintain a constant temperature. The control probe is a precision platinum resistance thermometer (PRT). It is delicate and must be handled carefully. The probe is placed in the small hole in the top of the bath so that the probe tip is fully immersed in the bath fluid. It is located underneath the motor cover.

The stirring motor is mounted on the bath tank lid under the motor cover. It drives the stirring propeller to provide mixing of the bath fluid. Proper mixing of the fluid is important for good constant temperature stability.

On the bath lid is a work area access hole. This is used for filling the bath with fluids and placement of thermometers and devices into the bath. When possible the access hole should be covered (must be covered to reach minimum temperatures).

An insulated access hole cover is provided and should be used to cover the access opening in the top of the bath. This improves bath temperature stability, prevents excess fluid evaporation or fumes and increases safety with hot fluid. The user may drill or cut holes in the cover to accommo- date the instruments to be calibrated or immersed in the bath. Spare cov- ers are available from GE Kaye Instruments. An optional access cover which provides locations for two reference thermometers and three wells for units under test is available.

7.3Back Panel

On the back of the bath are 1) the system fuses, 2) the non-removable power cord, 3) the drain valve, 4) the RS-232 interface connector, and 5) removable vent panel.

24

Image 28 Contents
GE Kaye World Headquarters Table of Contents General Operation Parts and ControlsController Operation Maintenance Trouble Shooting Digital Communication InterfaceWiring Diagram Figures Tables Symbols Used Before You StartSafety Information Canadian Standards AssociationBurn Hazard Bath Fluids Cold Baths Customer Service Information World HeadquartersIntroduction Specifications Environmental ConditionsHardware Warranty and Assistance Instrument WarrantyIn-Warranty Repairs After-Warranty RepairsEquipment Maintenance Agreements Customer SupportCustomer Support Agreement Customer Site AssistanceUnpacking Quick StartSet-up Power Setting the TemperatureAccess set-point selection Set-point 1, 25.0C currently used Installation Drain Valve Installation InstructionsBath Environment Dry-out PeriodFilling With Fluid Necessary Comparison Calibration GeneralBath Use Calibration of Multiple Probes Parts and Controls Control PanelBath Tank and Lid Back PanelBack Panel Temperature Range General OperationBath Fluid ViscositySafety Thermal ConductivityThermal Expansion Electrical ResistivityCost Commonly Used FluidsWater Distilled EthanolMineral Oil Fluid Characteristics TableSilicone Oil Dow Corning 200.05, 200.10 Limitations and Disclaimer Table of Various Bath FluidsHeater StirringRefrigeration Operation Temperature ControllerImportant Refrigerant Information General Operation Bath Temperature Temperature Set-pointController Operation Programmable Set-pointsDisplay Temperature Temperature Scale Units Set-point ValueScan Control ScanScan Rate Secondary Menu Heater PowerProportional Band New proportional band setting Cutout CO= 80CController Configuration Operating ParametersHigh Limit Low LimitSerial Interface Parameters CoolingBaud Rate Duplex ModeSample Period Calibration Parameters LinefeedHard Cutout 11.2 R0Alpha DeltaDigital Communication Interface Serial CommunicationsWiring SetupRS-232 Cable Wiring for IBM PC and CompatiblesInterface Commands Duplex ModeSerial Operation Sample PeriodSecondary Menu Lfeed=on Lf=on Calibration Procedure Calibration ProcedureCalibration Equipment CalibrationCalibration Procedure Maintenance Maintenance Trouble Shooting TroubleshootingCE Comments Low Voltage Directive SafetyEMC Directive Wiring Diagram Revisions