Ensure the valve handle is in the closed position before attempting to add fluid to the tank.

Hex Nut

Back of bath

Valve body

Figure 1 Drain Valve Installation—IMPORTANT: Do Not Over Tighten. Follow the installation instructions above.

5.3.2Filling With Fluid

The CTR –80 Bath is not provided with a fluid. Depending on the desired tem- perature range, any of the following fluids, as well as others, may be used in the bath:

Water

Ethanol (Ethyl Alcohol)

Ethylene glycol/water

Mineral oil

Silicone oil

Halocarbon 0.8

Fluids are discussed in detail in Section 8.1.

Remove any access hole cover from the bath and check the tank for foreign matter (dirt, remnant packing material, etc.).

Fill the bath with clean unpolluted fluid. Fill the bath carefully through the large square access hole to a level that will allow for stirring and thermal ex- pansion. DO NOT turn on the bath without fluid in the tank. The fluid should never exceed a height of 1/2" below the top of the tank or be less than 2 inches

18

Image 23 Contents
GE Kaye World Headquarters Table of Contents Controller Operation Parts and ControlsGeneral Operation Digital Communication Interface Maintenance Trouble ShootingWiring Diagram Figures Tables Before You Start Symbols UsedCanadian Standards Association Safety InformationBurn Hazard Bath Fluids Cold Baths World Headquarters Customer Service InformationIntroduction Environmental Conditions SpecificationsAfter-Warranty Repairs Hardware Warranty and AssistanceInstrument Warranty In-Warranty RepairsCustomer Site Assistance Equipment Maintenance AgreementsCustomer Support Customer Support Agreement Set-up Quick Start Unpacking Setting the Temperature PowerAccess set-point selection Set-point 1, 25.0C currently used Dry-out Period InstallationDrain Valve Installation Instructions Bath EnvironmentFilling With Fluid Necessary Bath Use GeneralComparison Calibration Calibration of Multiple Probes Control Panel Parts and ControlsBack Panel Bath Tank and LidBack Panel Viscosity Temperature RangeGeneral Operation Bath FluidElectrical Resistivity SafetyThermal Conductivity Thermal ExpansionEthanol CostCommonly Used Fluids Water DistilledSilicone Oil Dow Corning 200.05, 200.10 Fluid Characteristics TableMineral Oil Table of Various Bath Fluids Limitations and DisclaimerRefrigeration StirringHeater Important Refrigerant Information Temperature ControllerOperation General Operation Programmable Set-points Bath TemperatureTemperature Set-point Controller OperationDisplay Temperature Set-point Value Temperature Scale UnitsScan Rate ScanScan Control Proportional Band Heater PowerSecondary Menu New proportional band setting CO= 80C CutoutLow Limit Controller ConfigurationOperating Parameters High LimitCooling Serial Interface ParametersSample Period Duplex ModeBaud Rate Linefeed Calibration ParametersDelta Hard Cutout11.2 R0 AlphaSerial Communications Digital Communication InterfaceIBM PC and Compatibles WiringSetup RS-232 Cable Wiring forSample Period Interface CommandsDuplex Mode Serial OperationSecondary Menu Lfeed=on Lf=on Calibration Calibration ProcedureCalibration Procedure Calibration EquipmentCalibration Procedure Maintenance Maintenance Troubleshooting Trouble ShootingEMC Directive Low Voltage Directive SafetyCE Comments Revisions Wiring Diagram