(3.6) Others

(1) Contents and formats of text.

Item

Digit

Contents

Remarks

1.

Rack No.

4 or 5

'0001'-'9999' or

• The digit of rack no.

 

 

 

'00001'-'99999'

 

and cup pos. and

(Optional)

 

STAT sample is space

 

these using is

 

 

 

 

 

changeable in

 

 

 

When Rack no. is no setting,

 

[PARAMETER]

 

 

 

Rack No. is '00000'.

 

[ONLINE]

 

 

 

 

 

[PROTOCOL]

 

 

 

 

 

menu.

 

 

 

 

In case ,receive

 

 

 

 

 

T.R.I.

 

 

 

 

 

in batch.

2.

Cup position

2

rack : '01'-'10'

This is automatically

(Optional)

 

STAT : '01'-'22'

 

assigned when rack

 

 

 

 

 

No. is in use.

3.

Sample type

1

space : serum

 

 

 

 

 

U : urine

 

 

 

 

 

X : other

 

 

4.

Sample No.

4

'0001'-'9999' : Routine sample

 

 

 

 

 

'E001'-'E999' : Emergency sample

 

 

 

 

 

'P001'-'P999' : Stat sample

 

 

 

 

 

'R001'-'R999' : Reagent blank

 

 

 

 

 

'A001'-'A999' : Calibration sample

 

 

 

 

 

'Q001'-'Q999' : QC sample

 

 

5.

Sample ID

4-26

Number or character

• Number of digits can

 

 

 

 

 

be changed in the

 

 

 

 

 

[PARAMETER]

 

 

 

 

 

[FORMAT]

 

 

 

 

 

[FORMAT

 

 

 

 

 

REQUISITION]

 

 

 

 

 

menu.

6.

Original

4

'0001'-'9999' : normal sample

For Repeat Run Results.

 

Sample No.

 

'E001'-'E999' : emergency sample

 

 

 

 

 

'P001'-'P999' : stat sample

 

 

7.

Reserved

4

Space

 

 

8.

Block

1

'0'-'9' Last text is 'E'

 

 

 

Identification

 

 

 

 

 

No.

 

 

 

 

- 22 -

Page 22
Image 22
2Wire AU5400, AU2700 specifications Others, Requisition