9.ANL unit No.

1

‘1’: AU5400 ISE

ANL unit No. is added to the

 

 

‘1’: AU5400 unit 1

header portion of RB, ACAL,

(Optional)

 

‘2’: AU5400 unit 2

and QC result data. Normal

 

 

‘3’: AU5400 unit 3

sample and rerun sample, it is

 

 

 

added to each item. However,

 

 

‘1’: AU2700

in the case of an calculate

 

 

 

item, a space is added. In a

 

 

 

normal sample and repeat run,

 

 

 

it can be prevented from

 

 

 

adding a unit number by the

 

 

 

setup of an on-line condition

 

 

 

menu. This option can be

 

 

 

selected in the

 

 

 

[PARAMETER][ONLINE]

 

 

 

[PROTOCOL][SPECIAL]

 

 

 

menu.

10.Cuvette side

1

AU5400

Cuvette side is added to the

 

 

‘0’ : inner cuvette (ISE Cell 1)

header portion of RB, ACAL,

(Optional)

 

‘1’ : outer cuvette (ISE Cell 2)

and QC result data. Normal

 

 

AU2700

sample and rerun sample, it is

 

 

‘0’ : inner cuvette(ISE)

added to each item. However,

 

 

‘1’ : outer cuvette

in the case of an calculate

 

 

 

item, a space is added. In a

 

 

 

normal sample and rerun, it

 

 

 

can be prevented from adding

 

 

 

a unit number by the setup of

 

 

 

an on-line condition menu.

 

 

 

This option can be selected in

 

 

 

the

 

 

 

[PARAMETER][ONLINE]

 

 

 

[PROTOCOL][SPECIAL]

 

 

 

menu.

11. Sex

1

M : male

• These items can be

 

 

F : female

selected/canceled in

(Optional)

 

SP : none sex

the [PARAMETER]

 

 

0 : no-set

[FORMAT]

 

 

 

[FORMAT

 

 

 

REQUISITION]

 

 

 

menu.

12. Age

3

'000'-'150'

 

(Optional)

 

space : no-set *1

 

13. Month

2

'00'-'11'

 

(Optional)

 

space : no-set *1

 

14. Patient

20

number or character

 

Information

 

 

 

(Optional)

 

 

 

15. Online

2

'01'-'99'

• This number can be

Test No.

 

 

changed in the

 

 

 

[PARAMETER]

 

 

 

[ONLINE] [TEST

 

 

 

NO.]

16. Method

1

'0':normal

For Repeat Run Results.

 

 

'1':dilution

 

 

 

‘2’:condense

 

- 23 -