8)ONLINE ILLEGAL SAMPLE KIND No. [ aa ] <Operation>

(1).Disregard the test requisition information text for the sample generating the alarm.

(2).Continue to receive the test requisition information text for the other samples within the run. <Operation>

(1).When receiving the test requisition information text using online, sample kind No. of the sample generating the alarm does not meet the specification.

(2).The display of [ ] and its meaning are as follows.

aa: Sample type

----------------------------------------------------------------

__

: Normal sample

_E

: Emergency sample

_P

: Stat sample

U_

: Urine normal sample

UE

: Urine emergency sample

UP

: Urine stat sample

X_

: Other normal sample

XE

: Other emergency sample

XP

: Other stat sample

----------------------------------------------------------------

9)ONLINE ILLEGAL GENDER TEXT [ a ] <Operation>

(1).Disregard the test requisition information text for the sample generating the alarm.

(2).Continue to receive the test requisition information text for other samples within the run. <Description>

(1).When receiving the requisition information text using online, the sex of this text is not within the specification.

(2).The display of [ ] and its meaning are as follows.

a : Received sex 10)ONLINE ILLEGAL AGE/MONTH [ aaa bb ] <Operation>

(1).Disregard the test requisition information text for the sample generating the alarm.

(2).Continue to receive the test requisition information text for other samples within the run. <Description>

(1).When receiving the requisition information text using online, age or month age (age for baby before one year old) of this text is out of the specification.

(2).The display of [ ] and its meaning are as follows.

aaa

: Received age

bb

: Received month age

- 60 -

Page 60
Image 60
2Wire AU5400, AU2700 specifications