<Operation>

(1).Disregard the test requisition information text for the sample generating the alarm.

(2).Continue to receive the test requisition information text for the other samples within the run. <Description>

(1).When receiving the test requisition information text using online, sample type of this does not match the required sample.

(2).The display of [ ] and its meaning are as follows.

aaaaaaa: Transferred sample type and sample No.

---------------------------------------------------------------

__ nnnn : Normal sample

_E nnnn : Emergency sample

_P nnnn : Stat sample

U_ nnnn : Urine normal sample

UE nnnn : Urine emergency sample

UP nnnn : Urine stat sample

X_ nnnn : Other normal sample

XE nnnn : Other emergency sample

XP nnnn : Other Stat sample

---------------------------------------------------------------

bbbbbb : Received sample type and sample No.

---------------------------------------------------------------

__ nnnn : Normal sample

_E nnnn : Emergency sample

_P nnnn : Stat sample

U_ nnnn : Urine normal sample

UE nnnn : Urine emergency sample

UP nnnn : Urine stat sample

X_ nnnn : Other normal sample

XE nnnn : Other emergency sample

XP nnnn : Other Stat sample

---------------------------------------------------------------

- 62 -

Page 62
Image 62