3)ONLINE FORMAT ERROR <<Operation>>

(1).Disregard the test requisition information text for the sample generating the alarm.

(2).Continue to receive the test requisition information text for samples following the alarm. <<Description>>

(1).In the Online Parameter menu, submenu Set Up, the T.R.I. receive is not designated asn "YES" for the necessary parameter.

(2).The display in the [ ] and its meaning are as follows.

aaaa: Type of error

----------------------------------------------------------------

R_ __

: normal run, a normal sample

R_ _E : normal run, an emergency sample

R_ _P : normal run, a STAT sample

R_ U_

: normal run, a urine normal sample

R_ UE

: normal run, a urine emergency sample

R_ UP

: normal run, a urine STAT sample

R_ X_

: normal run, a Other normal sample

R_ XE

: normal run, an Other emergency sample

R_ XP

: normal run, a Other STAT sample

----------------------------------------------------------------

RH __

: repeat run, a normal sample

RH _E : repeat run, an emergency sample

RH _P : repeat run, a STAT sample

RH U_

: repeat run, a urine normal sample

RH UE

: repeat run, a urine emergency sample

RH UP

: repeat run, a urine STAT sample

RH X_

: repeat run, a Other normal sample

RH XE

: repeat run, an Other emergency sample

RH XP

: repeat run, a Other STAT sample

----------------------------------------------------------------

bbbbbbbbbb: Information type that is not designated No sample No.

No ID No.

- 56 -

Page 56
Image 56
2Wire AU5400, AU2700 specifications Rh U