2)ONLINE ERROR [ aa ] [ bbbbbbbbb ] <Operation>

(1). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "STOP", the following events can occur:

1). The text following the alarm will not be received.

2). In a case when the text requisition is received real-time,the "STOP" designation is ignored and the text requisition will be received in real-time at the start of the next analysis.

(2). In the Online Parameter menu, submenu Error Control, if the T.R.I. receive is designated as "Continue", the test requisition for the next sample will be received.

<Description>

(1). Communication error occurs when transmitting the text for test requisition or receiving the test requisition format.

(2). The display in the [ ] and its meaning are as follows.

aa: type of the errors

----------------------------------------------------------------

01 : Device name error

02 : Framing error

03 : Over run error

04 : Parity error

05 : Time out error

06 : Receive NAK when transmitting a text

07 : BCC error when receiving a text

08 : Other communication errors

09 : Function error (Full Duplex instead of Half Duplex)

10 : Unit name error

11 : Parameter error

12 : Request cancel

----------------------------------------------------------------

bbbbbbbbb : Text having error

----------------------------------------------------------------

RB

: Start text for receiving the test

requisition information.

RE

: End text for receiving the test

requisition information.

R_ __ nnnn

 

: normal run, normal sample, sample No.

R_ _E nnnn

 

: normal run, emergency sample, sample No.

R_ _P nnnn

 

: normal run, STAT sample, sample No.

R_ U_ nnnn

 

: normal run, urine sample, sample No.

R_ UE nnnn

 

: normal run, urine emergency sample, sample No.

R_ UP nnnn

 

: normal run, urine STAT sample, sample No.

R_ X_ nnnn

 

: normal run, Other normal sample, sample No.

R_ XE nnnn

 

: normal run, Other emergency sample, sample No.

R_ XP nnnn

 

: normal run, Other STAT sample, sample No.

RH __ nnnn

 

: repeat run, normal sample, sample No.

RH _E nnnn

 

: repeat run, emergency sample, sample No.

RH _P nnnn

 

: repeat run, STAT sample, sample No.

RH U_ nnnn

: repeat run, urine sample, sample No.

RH UE nnnn

: repeat run, urine emergency sample, sample No.

RH UP nnnn

: repeat run, urine STAT sample, sample No.

RH X_ nnnn

: repeat run, Other normal sample, sample No.

RH XE nnnn

: repeat run, Other emergency sample, sample No.

RH XP nnnn

: repeat run, Other STAT sample, sample No.

R_ __ xxxxxxxxxxx: normal run, normal sample, sample ID

R_ _E xxxxxxxxxxx: normal run, emergency sample, sample ID

R_ _P xxxxxxxxxxx: normal run, STAT sample, sample ID

R_ U_ xxxxxxxxxxx: normal run, urine sample, sample ID

R_ UE xxxxxxxxxxx: normal run, urine emergency sample, sample ID

R_ UP xxxxxxxxxxx: normal run, urine STAT sample, sample ID

R_ X_ xxxxxxxxxxx: normal run, Other normal sample, sample ID

R_ XE xxxxxxxxxxx: normal run, Other emergency sample, sample ID

R_ XP xxxxxxxxxxx: normal run, Other STAT sample, sample ID

RH __ xxxxxxxxxxx: repeat run, normal sample, sample ID

RH _E xxxxxxxxxxx: repeat run, emergency sample, sample ID

- 54 -

Page 54
Image 54
2Wire AU5400, AU2700 specifications