3M 390G Web Browser Requirements, A. Serial number URL, B. IP Address URL, Dynamic IP Address

Models: 390G

1 14
Download 14 pages 42.12 Kb
Page 10
Image 10
OPTIONAL 3M™ ATTEST™ AUTO‑READER 390G WEB APP

OPTIONAL 3M™ ATTEST™ AUTO‑READER 390G WEB APP

Web Browser Requirements

The Web Interface is designed to be viewed in Internet Explorer 8.0 or higher. Earlier versions of Internet Explorer or other browsers may result in some web pages being displayed in a less‑than‑optimum manner. For best results, switching to Internet Explorer

8.0or higher is recommended. For full functionality, the use of an HTML5 compliant browser with JavaScript enabled is recommended.

Connecting to the 3M™ Attest™ Auto‑reader 390G Web App

To use the 3M™ Attest™ Auto‑reader 390G Web App, the 3M™ Attest™ 390G Auto‑reader must be connected to the facility network using the supplied Ethernet cable. Consult with your facility IT personnel for assistance with network connectivity. Users may connect to the 3M™ Attest™ Auto‑reader 390G Web App using a URL based upon either the 3M™ Attest™ Auto‑reader 390G serial number or the IP Address. To view the 3M™ Attest™ Auto‑reader 390G Web App from a wireless device such as a mobile hand‑held device or a computer on a sub‑net different from that to which the 3M™ Attest™ Auto‑reader 390G is connected, use the IP Address URL.

Note: To access the 3M™ Attest™ Auto‑reader 390G Web App within a secure network, the PC or mobile device must have access to the secure network; options include connecting via a network Ethernet cable, signing in to the facility’s wireless network, or connecting via a virtual private network (VPN).

A. Serial number URL

1.Connect the supplied Ethernet cable from the 3M™ Attest™ Auto‑reader 390G to an Ethernet jack.

2.Depress the Web Browser Requirements button once to display the 3M™ Attest™ Auto‑reader 390G 6‑digit serial number on the left side of the top row of text on the LCD display.

3.Enter http://attestxxxxxx where xxxxxx represents the unit’s serial number (e.g., http:// attest401008) from Step 2 in PC web browser address bar and press Go.

4.3M™ Attest™ Auto‑reader 390G Web App Status page will display on PC. For future ease of use, add this page to your web browser “Favorites.”

B. IP Address URL

The 3M™ Attest™ Auto‑reader 390G Web App may also be accessed using either a dynamic or static IP address.

1. Dynamic IP Address

A dynamic IP address is assigned when the Ethernet cable is connected from the 3M™ Attest™ Auto‑reader 390G to a live Ethernet jack.

Press and release the Connecting to the 3M™ Attest™ Auto‑reader 390G Web App button twice to display the 3M™ Attest™ Auto‑reader 390G’s dynamic IP address. Enter http://IPaddress (e.g. http://123.45.678.90) in the web browser address bar of the PC or mobile hand‑held device and press Enter. The 3M™ Attest™ Auto‑reader 390G Web App Status page will display.

2. Static IP address

Some facility IT personnel may require devices connecting to the network to use a Static IP address. If this is the case at your facility, contact the IT department to request a Static IP address.

Record the static IP address provided by your facility IT department here: ___________.

You may wish to label the Ethernet jack with the assigned address. Enter http://StaticIPaddress (e.g. http://111.22.333.44) in the web browser address bar on your PC or mobile hand‑held device and press Enter. The 3M™ Attest™ Auto‑reader 390G Web App Status page will display.

About the 3M™ Attest™ Auto‑reader 390G Web App Pages

The 3M™ Attest™ Auto‑reader 390G Web App includes four pages:

a) Status page

 

 

 

 

 

 

3M™ Attest™ Auto-reader 390G

Status

 

 

 

 

 

 

 

 

 

History

 

 

 

 

 

 

 

 

 

 

ATTEST 1294

 

 

ATTEST 1294

 

1

2

3

4

5

6

7

8

9

10

+

85

240

 

 

 

 

 

 

 

 

 

 

2012-10-31 12:32

 

 

 

 

 

 

 

SN

 

100122

 

 

 

 

1.User‑entered nickname (optional)

2.Well Configuration (1‑5 and 6‑10)

3.Well Number

4.BI Result/Remaining Incubation Time

5.Date and Time

6.3M™ Attest™ Auto‑reader 390G Serial Number

The Status page provides a real‑time representation of the 3M™ Attest™

Auto‑reader 390G LCD panel display. The current date and time and 3M™ Attest™ Auto‑reader 390G serial number are also indicated on this page. From this page, the user may navigate to:

the History page by clicking on History on the left of the screen.

the Results Detail page by clicking on the well number of an occupied well (i.e., a well that currently has an incubating BI). Clicking on an unoccupied well takes the user to the History page.

b) Results Detail page

 

3M™ Attest™ Auto-reader 390G Results Detail

Status

 

History

 

Print

 

Result

Start Time

2012-10-29 11:08:04

Well# and Configuration

#3 1294

390G Serial Number

100122

Control?

 

BI Lot#

 

Load#

 

Sterilizer#

 

Technician

 

This page provides an individual record for an incubated BI. For complete documentation, the user may print a hard copy and/or save the record as an electronic file.

Note 1: The top four rows (highlighted in green) are populated automatically by the

3M™ Attest™ Auto‑reader 390G and cannot be changed by the user. The remaining rows are completed by the user to satisfy the documentation requirements of their facility.

Note 2: The Results Detail page refreshes less frequently that the Status page. Until the BI result is available, the Results Detail page displays remaining minutes of incubation in the results field.

1

9

Page 10
Image 10
3M 390G Web Browser Requirements, A. Serial number URL, B. IP Address URL, Dynamic IP Address, Static IP address