INSTRUCTIONS FOR USE – Wells, Buttons, & Display Interpretation
1
Please familiarize yourself with the face of the 3M™ Attest™ Auto‑reader 390G:
Wells
1.Incubation/Reader Wells – The 3M™ Attest™ Auto‑reader 390G has ten incubation wells.
Buttons
2. Information / Up Arrow Button
3. Temperature / Down Arrow Button
4. Buzzer Alarm OFF / Enter Button
Information Display
5.LCD Display – Display is backlit
6.LCD Display – Example of Error Message for well #6
7.LCD Display – Example of Remaining Incubation Time, in minutes, for well #10
8.LCD Display – Example of Negative ( ‑ ) BI result for well #9
9.LCD Display – Example of Positive ( + ) BI result for well #8
10.LCD Display – Alarm Enabled Indicator
Color‑Coded Configuration Indicator
11. Incubation well configuration sticker is color‑coded green
BI Crusher
12. BI Crusher Well
INSTRUCTIONS FOR USE – 3M™ Attest™
Auto‑reader 390G Configuration
Note: Ideally, the 3M™ Attest™ Auto‑reader 390G should be empty (i.e., have no incubating BIs) when it is reconfigured.
Press the and buttons on the front of the 3M™ Attest™ Auto‑reader 390G simultaneously for 3 seconds to enter configuration mode. Once in configuration mode,
use the Up and Down buttons to set the 3M™ Attest™ Auto‑reader 390G
clock and audible alarm as desired. The button acts as an “Enter” key and moves the flashing cursor to the next changeable setting.
Setting the 3M™ Attest™ Auto‑reader 390G Clock
The 3M™ Attest™ Auto‑reader 390G date and time are displayed in this format:
YYYY‑MM‑DD HH:MM, with the time displayed using a 24 hour clock. To change the year,
if necessary, push the up or down buttons to set desired year, then press to advance
to the month, adjust as necessary and press to advance to the day. Use the button to move to the time and set the hours and minutes.
Setting the Audible Alarm Function
Move across the top row of text by pushing the button. The flashing “B” in Buzzer indicates this is a changeable setting. It is recommended this be set at “Buzzer Enabled” to audibly alert the operator to a positive BI but it may be set to “Buzzer Disabled” by using either the up or down button. Note: If Buzzer Disabled is selected, and the changes accepted, the speaker icon ( ) will disappear from the LCD panel and the alarm will be disabled for all positive 3M™ Attest™ Rapid Readout Biological Indicators, error codes, and caution code alarm warnings. Note: The 3M™ Attest™ Auto‑reader 390G display backlight flashes whenever an audible alarm sounds. Disabling the audible alarm does NOT disable the LCD display backlight flashing.
Press to select the desired option and advance to the Accept screen.
Accept Screen
Use the up or down buttons to toggle between Yes and No, press to accept changes (Yes selected) or retain current settings (No selected) and exit the configuration mode. The LCD panel will return to the standard display.
5