1
SPECIFICATIONS
Power Supply Specifications
Power supply | Operating Condition | Units |
Voltage Range | 100‑240 | Volts AC |
Frequency | 50/60 | Hertz |
Current | 0.6 | Amps |
Environmental Operating Conditions
Environmental Condition | Operating Condition | Units |
Altitude | 3000 (max) | Meters |
Operating Temperature | 16 – 29 | ºC |
Storage Temperature | ‑10 to 50 | ºC |
Operating Relative Humidity | 20‑80 (non‑condensing) | % |
Voltage Range | 12 | Volts DC |
Frequency | DC |
|
Current | 1.2 | Amps |
Installation/Over Voltage | Category II |
|
Pollution Degree | 2 |
|
INSTRUCTIONS FOR USE – Initial Set‑up
Choose Location
Place 3M™ Attest™ Auto‑reader 390G on a firm and level surface, away from sunlight and incandescent light. To save counter space, units may be stacked two high.
Select & Install Correct Power Plug
The 3M™ Attest™ Auto‑reader 390G power supply comes with multiple blade configurations. Install the blade configuration appropriate for your country into the power supply and plug into appropriate AC grounded outlet.
WARNING:
•Do not place the instrument in environment exposed to sunlight or strong incandescent light.
•Do not place instrument close to any device that emits a strong electro‑magnetic field.
•Do not use on a vibrating surface.
WARNING: Use only the power supply specified for this product and certified for the country of use.
Wait 30 Minutes Before Using
Once plugged in, allow a 30 minute warm‑up period before placing 3M™ Attest™ Rapid Readout Biological Indicators into the incubation wells. The C1 caution code will disappear when proper incubation temperature is reached.
Special Considerations
It is recommended that the unit be left on to eliminate warm‑up periods. The unit may be turned off when it is anticipated that it will not be used for an extended period.