19

Views on the Edge

Definitions of data log values and views

Definitions of data log values and views

To set up the Edge and work with noise-level data, you will want to have a good understanding of the terms listed in Table 3-1.

NOTE: The following table is based on A-weighted, slow measurements. Thus, if you choose A or C-weighted with Fast or Slow response the measurement values will change on the display (i.e., you selected sound pressure level with C-weighting with Fast response, it will display as LCF.) For further details, please see “Dosimeter measurement descriptors/variants” in Table 3-2.

 

 

 

 

 

 

Views on the

 

Explanation

 

 

EDGE

 

 

 

 

LAS

 

Sound pressure level - is the basic measure of noise loudness,

 

 

 

expressed in decibels. (In DMS this is noted as “SPL”)

 

 

 

 

 

 

LASMX

 

Maximum sound level- is the highest sound level recorded during a

 

 

 

measurement interval with a particular response setting (Slow or Fast) and

 

 

 

weighting (A or C). LMax is the highest value that is included in LAvg or

 

 

 

dose calculations.

 

 

 

 

Under OSHA regulations, unprotected workers may not be exposed to

 

 

 

maximum sound levels greater than 115 dB, measured with a Slow

 

 

 

response rate.

 

 

 

 

 

 

LCPK

 

Peak Level - is the highest instantaneous sound pressure level recorded

 

 

 

during a measurement interval. Unlike the maximum level, the peak is

 

 

 

detected independently of dosimeter settings for response rate or

 

 

 

weighting.

 

 

 

 

Under OSHA regulations, unprotected workers may not be exposed to

 

 

 

peak sound levels greater than 140 dB.

 

 

 

 

 

 

LASMN

 

Minimum sound level - The lowest sampled sound level during the

 

 

 

measurement interval allowing for the unit’s Response Time setting (Fast

 

 

 

or Slow).

 

 

 

 

 

 

LAVG

 

Average sound level - (in dB) computed for a chosen time duration,

 

 

 

based on a 4, 5, or 6 dB exchange rate. LAVG with a 3 dB exchange rate is

 

 

 

referred to as LEQ (equivalent sound level). All sound levels at or above

 

 

 

the threshold level are averaged into the calculations relating to noise

 

 

 

exposure. LAVG is typically calculated with no threshold, with a threshold of

 

 

 

80 dB or with a threshold of 90 dB.

 

 

 

Table 3-1: Dosimeter views and definitions

 

053-644, Rev H

 

Page 19

eg4

Page 25
Image 25
3M eg4 user manual Definitions of data log values and views, Views on Explanation, Dosimeter views and definitions