MAINTENANCE
INSPECTION AND CLEANING
Perform the following inspection and cleaning procedures periodically as a preventive maintenance measure to keep the Acclaim Camera System in optimal condition resulting in
DO NOT ATTEMPT ANY INTERNAL SERVICE OF ACCLAIM CAMERA SYSTEM COMPONENTS.
CONTACT YOUR LOCAL AUTHORIZED AIR TECHNIQUES DEALER FOR SERVICE. FAILURE TO
HEED THIS WARNING MAY RESULT IN EQUIPMENT DAMAGE AND VOIDING THE WARRANTY.
Inspection - Routinely inspect each component of the Acclaim Camera System for possible defects as follows:
1.Camera Handpiece -
a.Check overall handpiece for chips, cracks or other irregularities.
b.Check the lens window for debris or spots.
c.Check the connector socket for damage.
2.Control Module -
a.Check for broken indicator lamp and switch.
b.Check all connector sockets for damage, loose or missing hardware.
3.Cables and Connectors -
a.Check cables for damaged or deteriorated insulation kinking or twisting.
b.Check connectors for loose, bent or missing pins.
c.Check that both
CAUTION: TURN OFF POWER AND UNPLUG THE ACCLAIM CAMERA SYSTEM COMPONENTS BEFORE PERFORMING CLEANING PROCEDURES.
DO NOT SPRAY CLEANING LIQUIDS OR DISINFECTANTS DIRECTLY
ON THE CAMERA HANDPIECE OR CONTROL MODULE. USE CARE
NOT TO ALLOW LIQUIDS TO RUN INTO INTERNAL CIRCUITRY.
DO NOT WIPE THE SURFACES USING BENZINE, THINNER, ETC. AS
THIS MAY DEGRADE THE FINISH.
Cleaning - Clean each component of the Acclaim Camera System as necessary by performing the following procedures.
1.Control Module - Use a
2.Camera Handpiece - Following every patient use, the Camera Handpiece can be wiped clean and disinfected with any
3.Camera Window Lens - Remove debris or spots from lens window by cleaning with alcohol and cotton swabs.
21