c.Check the middle ohms range as follows:

1. Press

, then

60501B:.

60502B:.

2.Set power supply for: 60501B: 10.9V and 8A.

60502B: 10.9V and 15A.

3.Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the Electronic Load resistance. The result should be between:

60501B: 46 and 86 ohms.

60502B: 24.1 and 39.6 ohms.

4.Then press:

60501B: Manual background.

60502B: Manual background.

5.Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the Electronic Load resistance. The result should be between:

60501B: 1.97 and 2.03 ohms.

60502B: 0.989 and 1.011 ohms.

d.Check the high ohms range as follows:

1. Press:

, then

60501B:.

60502B:

2.Set power source for: 60501B: 60V and 5A.

60502B: 60V and 6A.

3.Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the Electronic Load resistance. Calculation should be between:

60501B: 99.8 and 10K ohms.

60502B: 61.1 and 3243 ohms.

4.Then press:

60501B: Manual background.

60502B: Manual background.

5.Measure the voltage across the monitor resistor and across the module's input terminals, then calculate the Electronic Load resistance. The result should be between:

60501B: 21.4 and 27.3 ohms.

60502B: 10.9 and 13.3 ohms.