Verification

2

Verification

Introduction

This chapter contains test procedures that check the operation and calibration of the Agilent 60501B/60502B Electronic Load Modules. The tests are performed from the front panel of an Agilent 6050A/6051A Electronic Load Mainframe with the module installed in slot 1. The tests can also be used to determine which circuits are faulty when troubleshooting. There are some transient, trigger, and pulse functions that require an GPIB controller and will not be verified with manual testing from the front panel. The following tests will verify, with a high level of confidence, that the module is operating properly without testing all of its capabilities.

At the end of this chapter are performance record tables where actual measured values can be recorded.

Test Equipment Required

Table 2-1 lists the test equipment required to perform the tests in this chapter. Test setups for the tests are shown in Figures 2-1 through 2-3. Make sure the sense switch on the rear of the module is set to the LCL position since local sensing is used in all of the test setups. Use adequate wire gauge when making connections (see Chapter 3 in the Operating Manual).

 

Note

The Electronic Load must pass the selftest at power turn-on before the following tests can be performed.

 

 

 

If the unit fails selftest, refer to the overall troubleshooting procedures in the mainframe Service Manual.

 

 

 

 

Table 2-1. Test Equipment Required for Verification

 

 

 

 

 

 

 

 

Type

 

Required Characteristics

Recommended Model

 

 

60V/60A Source

 

0 to 20V/0 to 120A

Agilent 6031A or equivalent

 

 

 

 

0 to 60V/0 to 50A

Agilent 6032A or equivalent

 

 

Current Monitor

 

0.10 ohms @ 15A

Guideline 9230/15

 

 

Resistor

 

0.04% @ 25W

 

 

 

Current Monitor

 

0.010 ohms @ 100A

Guideline 9230/100

 

 

Resistor

 

0.04% @ 100W

 

 

 

Digital Voltmeter

 

dc accuracy of 0.01%

Agilent 3455A, 3456A, or 3458A

 

 

 

 

6 digit readout

 

 

 

Current Probe with

Sensitivity of 1mA/10 mV to 50MHz with

Tektronix A6302 probe, AM503 probe amplifier,

 

 

Amplifier and Power

less than 300∝A of noise to 5MHz.

and TM501 probe power supply.

 

 

Supply

 

 

 

 

 

Oscilloscope

 

Sensitivity: 1mV

Agilent 54504

 

 

 

 

Bandwidth: 20MHz

 

 

 

 

 

 

 

 

7