Verification
Introduction
This chapter contains test procedures that check the operation and calibration of the Agilent 60501B/60502B Electronic Load Modules. The tests are performed from the front panel of an Agilent 6050A/6051A Electronic Load Mainframe with the module installed in slot 1. The tests can also be used to determine which circuits are faulty when troubleshooting. There are some transient, trigger, and pulse functions that require an GPIB controller and will not be verified with manual testing from the front panel. The following tests will verify, with a high level of confidence, that the module is operating properly without testing all of its capabilities.
At the end of this chapter are performance record tables where actual measured values can be recorded.
Test Equipment Required
Table
| Note | The Electronic Load must pass the selftest at power |
| ||
|
| If the unit fails selftest, refer to the overall troubleshooting procedures in the mainframe Service Manual. |
| ||
|
|
| Table |
| |
|
|
|
|
|
|
| Type |
| Required Characteristics | Recommended Model |
|
| 60V/60A Source |
| 0 to 20V/0 to 120A | Agilent 6031A or equivalent |
|
|
|
| 0 to 60V/0 to 50A | Agilent 6032A or equivalent |
|
| Current Monitor |
| 0.10 ohms @ 15A | Guideline 9230/15 |
|
| Resistor |
| 0.04% @ 25W |
|
|
| Current Monitor |
| 0.010 ohms @ 100A | Guideline 9230/100 |
|
| Resistor |
| 0.04% @ 100W |
|
|
| Digital Voltmeter |
| dc accuracy of 0.01% | Agilent 3455A, 3456A, or 3458A |
|
|
|
| 6 digit readout |
|
|
| Current Probe with | Sensitivity of 1mA/10 mV to 50MHz with | Tektronix A6302 probe, AM503 probe amplifier, |
| |
| Amplifier and Power | less than 300∝A of noise to 5MHz. | and TM501 probe power supply. |
| |
| Supply |
|
|
|
|
| Oscilloscope |
| Sensitivity: 1mV | Agilent 54504 |
|
|
|
| Bandwidth: 20MHz |
|
|
|
|
|
|
|
|