Agilent Technologies 6811B, 6813B, 6812B manual Calibrating and Entering DC Voltage Gain Values

Models: 6811B 6813B

1 91
Download 91 pages 25.18 Kb
Page 73
Image 73

Verification and Calibration - B

11.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the dc voltage value displayed on the DVM.

 

12.

Press Shift Calibration, scroll to the CAL LEV P1 command, use ° to

CAL:LEV P4

 

scroll to the P4 parameter, and press Enter.

 

13.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the dc voltage value displayed on the DVM.

 

The ac source is now holding the new voltage offset calibration constants in RAM.

Calibrating and Entering DC Voltage Gain Values

Connect the negative terminal of the DVM to the COM output terminal. Note that some calibration points are negative; be sure to enter the CAL:DATA as negative values when required.

Action

Display

14.Connect the DVM (dc volts mode) directly to the ac source. Do not connect the ratio transformer or load resistor shown in Figure B-1.

15.

Press Shift Calibration, scroll to the CAL VOLT DC command, and

CAL:VOLT:DC

 

press Enter.

 

16.

Press Shift Calibration, scroll to the CAL LEV P1 command, and press

CAL:LEV P1

 

Enter to select the first calibration point.

 

17.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the dc voltage value displayed on the DVM.

 

18.

Press Shift Calibration, scroll to the CAL LEV P1 command, use ° to

CAL:LEV P2

 

scroll to the P2 parameter, and press Enter. This selects the second

 

 

calibration point.

 

19.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the dc voltage value displayed on the DVM.

 

20.

Press Shift Calibration, scroll to the CAL LEV P1 command, use ° to

CAL:LEV P3

 

scroll to the P3 parameter, and press Enter.

 

21.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the dc voltage value displayed on the DVM.

 

The ac source is now holding the new dc voltage gain calibration constants in RAM.

Calibrating and Entering AC rms Voltage Gain Values

Action

Display

22.Connect the DVM (ac volts mode) to the ac source using the ratio transformer shown in Figure B-1.

23.

Press Shift Calibration, scroll to the CAL VOLT AC command, and

CAL:VOLT:AC

 

press Enter.

 

24.

Press Shift Calibration, scroll to the CAL LEV P1 command, and press

CAL:LEV P1

 

Enter to select the first calibration point.

 

25.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the product of the transformer ratio times the ac voltage

 

 

value displayed on the DVM.

 

73

Page 73
Image 73
Agilent Technologies 6811B, 6813B, 6812B Calibrating and Entering DC Voltage Gain Values, Verification and Calibration - B