Agilent Technologies 6812B, 6813B Close the Unit, Figure D-1. Line Voltage Conversion Components

Models: 6811B 6813B

1 91
Download 91 pages 25.18 Kb
Page 86
Image 86
Close the Unit

D - Line Voltage Conversion

You need only to lift the board out of the way, you do not need to disconnect any cables.

Locate the connector in front of the power transformer.

For 100 V, 200 V, or 208 V operation, install the plug on the connector labeled: 200 V input.

For 120 V or 230 V operation, install the plug on the connector labeled: 240 V input.

Replace the pc board and inner cover.

NOTE: Be sure to replace all of the screws removed in steps 9 and 11.

Close the Unit

Replace the outer cover.

Change the label on the rear panel to reflect the changed input voltage rating.

Reconnect the power and turn on the unit.

NOTE: Line voltage conversion does not require any fuse changes.

RIGHT SIDE

(withcover removed)

200/220 100/120Figure D-1. Line Voltage Conversion ComponentsManual backgroundManual background

!! WARNING !!

If LEDs areon,

HAZARDOUS

voltagesarepresent!

Jumper Settings

 

100

J353

120

200

 

 

230

LEFT SIDE

(with cover removed)

200V INPUT

240V INPUT

JUMPER WIRE

LINE JUMPERS

POWER TRANSFORMER

Figure D-1. Line Voltage Conversion Components

86

Page 86
Image 86
Agilent Technologies 6812B, 6813B, 6811B manual Close the Unit, Figure D-1. Line Voltage Conversion Components