Agilent Technologies 6812B, 6813B, 6811B manual Calibrating the OVP Trip Point

Models: 6811B 6813B

1 91
Download 91 pages 25.18 Kb
Page 74
Image 74

B - Verification and Calibration

26.

Press Shift Calibration, scroll to the CAL LEV P1 command, use ° to

CAL:LEV P2

 

scroll to the P2 parameter, and press Enter. This selects the second

 

 

calibration point.

 

27.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the product of the transformer ratio times the ac voltage

 

 

value displayed on the DVM.

 

28.

Press Shift Calibration, scroll to the CAL LEV P1 command, use ° to

CAL:LEV P3

 

scroll to the P3 parameter, and press Enter.

 

29.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the product of the transformer ratio times the ac voltage

 

 

value displayed on the DVM.

 

30.

Press Shift Calibration, scroll to the CAL LEV P1 command, use ° to

CAL:LEV P4

 

scroll to the P4 parameter, and press Enter.

 

31.

Press Shift Calibration, scroll to CAL DATA 0.00, and use the Entry

CAL:DATA 0.00

 

keypad to enter the product of the transformer ratio times the ac voltage

 

 

value displayed on the DVM.

 

The ac source is now holding the new ac rms voltage calibration constants in RAM.

Calibrating the OVP Trip Point

 

Action

Display

32.

Press Shift Calibration, scroll to CAL VOLT PROT, and press Enter.

CAL:VOLT:PROT

33.

Wait for the ac source to compute the OVP calibration constant. The

 

 

display returns to Meter mode when the OVP calculation is complete.

 

The ac source is now holding the new OVP calibration constants in RAM.

Calibrating and Entering rms Current Values

Action

Display

34.Connect the DVM (ac rms mode), current shunt, and load resistor as shown in figure B-1 with S1 closed.

35.

Press Shift Calibration, scroll to the CAL CURR AC command, and

CAL:CURR:AC

 

press Enter.

 

36.

Press Shift Calibration, scroll to the CAL LEV P1 command, and press

CAL:LEV P1

 

Enter.

 

37.

Press Shift Calibration, scroll to the CAL DATA 0.00 command.

CAL:DATA 0.00

 

Calculate the rms current value (DVM ac rms voltage/shunt resistance), and

 

 

use the Entry keypad to enter the rms current.

 

38.

Press Shift Calibration, scroll to the CAL LEV P1 command, use ° to

CAL:LEV P2

 

scroll to the P2 parameter, and press Enter.

 

39.

Press Shift Calibration, scroll to the CAL DATA 0.00 command.

CAL:DATA 0.00

 

Calculate the rms current value (DVM ac rms voltage/shunt resistance), and

 

 

use the Entry keypad to enter the rms current.

 

The ac source is now holding the new rms current calibration constants in RAM.

74

Page 74
Image 74
Agilent Technologies 6812B, 6813B, 6811B manual Calibrating the OVP Trip Point, Calibrating and Entering rms Current Values