Examples of Front Panel Programming

Front Panel Operation - 4

Examples of Front Panel Programming

You will find these examples on the following pages:

1Setting the output voltage amplitude

2Setting the output frequency

3Setting the dc offset

4Setting a protection feature

5Clearing a protection feature

6Generating step, pulse, and list transients

7Programming trigger delays and phase synchronization

8Programming slew rates

9Measuring peak inrush current

10Setting the GPIB address or RS-232 parameters

11Saving and recalling operating states

The examples in the ac source Programming Guide are similar to the ones in this section, except that they use SCPI commands.

1 - Setting the Output Voltage Amplitude

NOTE: The maximum voltage that the ac source can output is limited by the peak value of the waveform, which is 425 Vpeak. Since the output is programmed in units of rms volts, the maximum value that can be programmed is dependent on the peak-to-rms ratio of the selected waveform. For a sinewave, the maximum ac voltage that can be programmed is 300 Vrms. For other waveforms the maximum may be different.

When you turn on the ac source, the default output shape is a 60 Hz sinewave at 0 Vrms. There is no output from the ac source because the default output state is OFF, as indicated by the Dis annunciator.

Set the output to 120 V rms as follows:

Action

You can set the voltage in any of three ways:

1.On the Function keypad, press Voltage. On the Entry keypad, press 1 2 0 Enter. This is the easiest way to enter an accurate value.

Display

VOLT 120

2.

On the Function keypad, press Voltage. On the Entry keypad, press or ° to

VOLT

127

 

increment or decrement the existing value. This technique is useful when you are

 

 

 

making minor changes to an existing value.

 

 

3

Rotate the front panel Voltage knob to obtain 120 V. This method is best when you

120 V

60 Hz

 

want to enter a value without using the voltage menu.

 

 

 

Note: You will not see the new voltage on the front panel meter unless the output is

 

 

 

enabled.

 

 

To enable the output:

 

 

4.

On the Function keypad, press Output On/Off. The Dis annunciator will go off,

120 V

60 Hz

 

indicating that the voltage is now applied to the output terminals.

 

 

47

Page 47
Image 47
Agilent Technologies 6812B, 6813B, 6811B manual Examples of Front Panel Programming, Setting the Output Voltage Amplitude