Programming output transients

Programming output transients

Up to now the ac source has been programmed with the transient system in Fixed mode. The following examples briefly describe the transient system’s Step, Pulse, and List modes, which require the application of a trigger to implement the transient mode.

NOTE: For the examples that follow, press Shift Output, scroll to *RST and press Enter to reset the unit prior to each example. Also press Enter to enter or activate each selection.

Program an output step

Voltage

VOLT:M STEP VOLT 120 VOLT:T 150

Trigger Control

INIT IMMED

Shift

 

Trigger

 

 

 

Step transients transition to a new output level upon receipt of a trigger. When these commands are sent, the voltage amplitude is stepped from its previous setting to 150 V rms upon receipt of a trigger.

TRIG level

VOLT level

Trigger

Program an output pulse

Voltage

VOLT:M PULSE VOLT 120 VOLT:T 90

Pulse

WIDTH .01

PER .03

COUNT 2

Trigger Control

INIT IMMED

Shift

 

Trigger

 

 

 

Pulse transients transition to a new output level upon receipt of a trigger and return to the original level after a specified time, repeating this action by the number of times specified by the count. When these commands are sent, two output pulses step the voltage amplitude from its previous setting to 90 V rms upon receipt of a trigger. At the end of the specified period (multiplied by the count), the voltage returns to its original level.

VOLT level

TRIG level

Width

Period

Trigger

Pulse Count

Program an output list

Voltage

VOLT: M LIST VOLT 120

Shift

 

List

 

 

 

DWELL [0] .5 DWELL [1] .5 DWELL [2] .5 VOLT [0] 130 VOLT [1] 140 VOLT [2] 150 STEP AUTO

Trigger Control

INIT IMMED

Shift

 

Trigger

 

 

 

List transients generate complex output sequences. When these commands are sent, the voltage amplitude is sequentially stepped to three levels upon receipt of a trigger, and then returns to the original voltage level. The output remains at each list step for .5 seconds. The values inside the brackets ([ ]) are the list index references. Use Clear Entry to clear a list.

Step 2

Step 1

Step 0

VOLT level

List

Trigger

complete

13