3-4
In Case of Difficulty
If the Plug-in Does Not Operate
If the Plug-in Does Not Operate If the plug-in does not operate:
1Make the following checks: 
• Is the plug-in module firmly seated in the ma inframe slot? 
• Are the knurled screws at the bottom of the plug-in module finger-tight? 
• If other equipment, cables, and connectors are be ing used with the plug-in 
module, are they connected properly and operating correctly?  
• Review the procedure for the test being performed when the problem ap-
peared. Are all the settings correct? Can  the problem be reproduced? 
• Are the connectors clean? Refer to the Agilent 83480A, 54750A User’s 
Guide for more information about cleaning the connectors. 
• Does the plug-in module work in the other two slots (1,2 or 3,4)? 
2Perform the following procedures: 
• Make sure the instrument is ready to acquire data by pressing Run. 
• Verify the pulse output is connected to channel 1 on the Agilent 83483 A, 
54751A electrical plug-in module. 
• Verify the trigger output is connected to the trigger input. 
• Find any signals on the channel inputs by pressing Autoscale. 
• Verify the Agilent 8133A pulse generator is set to: 
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . square wave 
Amplitude. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 mV 
Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 MHz 
Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 mV 
Pulse/trigger. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . enabled 
• See if any signals are present at the channel inputs by pressing: 
Trigger, Sweep, freerun