mode. In this mode the voltage of the negative output can be set lower than that of the positive output. The tracking ratio control allows the negative supply's output to be set to any value between a maximum that is within ±5% of the positive supply's output and a minimum that is less than 0.5 volts. Once a ratio is established by the tracking ratio control, the ratio of the positive output voltage to the negative output voltage remains constant as the ±20V voltage control varies the 0 to +20V output over its range.
The front panel also contains a line switch, three overload indicators for +6V output, +20V output, and
are available for nominal inputs of 100 and 230 Vac. The sup- ply is furnished with a detachable
LINE FUSE
The line fuse is located by the ac line receptacle. Check the rating of the line fuse and replace it with the correct fuse if necessary as indicated below. These are
Line Voltage | Fuse | Agilent Part No. |
100/115 Vac | 1.6 A | |
230 Vac | 1.0 A |
SPECIFICATIONS
Table 1 lists detailed specifications for the power supply.
Table 1. Specifications
AC INPUT
Standard: | 115 Vac ± 10%, |
OE9: | 100 Vac ± 10%, |
OE3: | 230 Vac ± 10%, |
DC OUTPUT and OVERLOAD PROTECTION
0 to ±20 V Outputs: Maximum rated output current is 0.5 A. Short circuit output current is 0.55 A ± 5% and a fixed cur- rent limit circuit limits the output of each supply to this maxi- mum at any output voltage setting. Unbalanced loads within current rating are permitted.
OPERATING TEMPERATURE RANGE
0 to 40 oC for full rated output. At higher temperatures, out-
put current is derated linearly to 50% at 55 oC maximum temperature.
TEMPERATURE COEFFICIENT
All Outputs: Less than 0.02% plus 1 mV voltage change
per oC over the operating range from 0 to 40 oC after 30 minutes
STABILITY (OUTPUT DRIFT)
0 to +6 V Output: Maximum rated output current is 2.5 A at 6 V. The maximum available output current decreases with the output voltage setting. A current foldback circuit limits the output to 2.75 A ± 5% at 6 volts and, with decreasing voltage, reduces the current limit linearly to 1 A ± 15% at zero volts (short circuited).
TRACKING ACCURACY
The +20 V and
LOAD REGULATION
All Outputs: Less than 0.01% plus 2 mV for a full load to no load change in output current.
LINE REGULATION
All Outputs: Less than 0.01% plus 2 mV for any line volt- age change within rating.
PARD (Ripple and Noise)
All Outputs: Less than 0.35 mV rms/1.5 mV
Common Mode Current (CMI): Less than 1 mA for all
outputs (20
All Outputs: Less than 0.1% plus 5 mV (dc to 20 Hz) dur- ing 8 hours at constant line, load and ambient after an initial
LOAD TRANSIENT RESPONSE TIME
All Outputs: Less than 50 msec for output recovery to within 15 mV of nominal output voltage following a load change from full load to half load, or vice versa.
OUTPUT VOLTAGE OVERSHOOT
All Outputs: During
METER ACCURACY: ±(0.5% of output + 2 counts) at
25oC±5oC
METER RESOLUTION
All Output: Voltage 10 mV Current 10 mA
DIMENSIONS
212.3mmW x 88.1 mmH x 269.2 mmD
(8.354 inW x 3.469 inH x 10.591 inD)
WEIGHT: 3.8 kg(8.4 lbs) net, 5.1 kg(11.3 lbs) shipping