44 Chapter 2
Agilent E4408B Specifications and Characteristics
Frequency
Agilent E4408B Specifications and
Characteristics
Specifications Supplemental Informati on
Sweep Time
Range 4 ms to 4000 sa
Tracking Generator On
(Option 1DN) 50 ms is the minimum sweep
time
Accuracy
(Span = 0 Hz)
4 ms to 4000 sa±1%
Sweep Triggerb Free Run, Singl e, L ine , V ideo c,
External, Del ayed, Offsetd
Delayed Trigger,e
Range 1 µs to 400 s
Resolution
rounded up to nearest µs
Accuracy ±(500 ns + (0.01% of delay))
Offset Triggerd
Resolution
Range ±327 ms to ±323 ks Where ST = sweep time
a. For firmware revisions prior to A.04.00, 5 ms to 2000 s.
b. Auto align is suspe nde d in vide o, ext ernal , and d elaye d trigg er m odes wh ile waiting for a
trigger event to occur.
c. Unavailable when RBW 300 Hz (Option 1DR).
d. For firmwa r e re v ision A.04.00 or late r.
e. Delayed trigger is available with line and external trigger.
Specifications Supplemental Informati on
Sweep (trace) Points 401
delay in seconds
65000
----------------------------------- ----
sweep time
400
---------------------------
32766 ST×400
------------------------------- to 32365 ST×
400
----------------------------