50 Chapter 2
Agilent E4408B Specifications and Characteristics
Amplitude
Agilent E4408B Specifications and
Characteristics
Specifications Supplemental Informati on
Marker Readout
Resolution
Log scale
RBW 1 kHz
0 to 85 dB from ref level 0.04 dB
RBW 300 Hz (Option 1DR)
0 to 120 dB from ref level 0.04 dB
Linear scale 0.01% of Reference Level
Specifications Supplemental Informati on
Frequency Response
Absolutea/Relative
10 dB attenuation
9 kHz to 3.0 GHz
20 to 30 °C±0.5 dB
0 to 55 °C±1.0 dB
Preselector centered for
frequency >3.0 GHz
10 dB attenuation
3.0 GHz to 6.7 GHz
Absolutea
20 to 30 °C ±1.5 dB
0 to 55 °C±2.5 dB
Relative
20 to 30 °C±1.3 dB
0 to 55 °C ±1.5 dB
6.7 GHz to 13.2 GHz
Absolutea
20 to 30 °C±2.0 dB
0 to 55 °C±3.0 dB