Frequency Span
Range 0 Hz (zero span), 100 Hz to
1.5 GHz
Resolution 2 Hz
Accuracy ±1.0% of span
Specifications Supplemental Information
Sweep Time
Range 4 ms to 4000 sa
Tracking Generator On
(Option 1DN or 1DQ) 50 ms is the minimum sweep
time
Accuracy
(Span = 0 Hz)
4 ms to 4000 sa±1%
Sweep Triggerb Free Run, Single, Line, Videoc,
External, Delayed, Offsetd
Delayed Trigger,e
Range 1 µs to 400 s
Resolution
rounded up to nearest µs
Accuracy ±(500 ns + (0.01% of delay))
Offset Triggerd
Resolution
Range ±327 ms to ±323 ks Where ST = sweep time
a. For firmware revisions prior to A.04.00, 5 ms to 2000 s.
b. Auto align is sus p end e d in vid eo, extern al, and delayed trig ge r mo d es whil e waiting for a
trigger event to occur.
c. Unavailable when RBW 300 Hz (Opt ion 1DR).
d. For firmware revision A.04.0 0 or later.
e. Delayed trigger is available with line and external trigger.
delay in seconds
65000
----------------------------------- ----
sweep time
400
---------------------------
32766 ST×400
------------------------------- to 32365 ST×
400
----------------------------