58 Chapter 2
Agilent E4408B Specifications and Characteristics
Options
Agilent E4408B Specifications and
Characteristics
Options
Tracking Generator (Option 1DN)

The spectrum analyzer/tracking generator combination will meet its

specification after a cable (812 0-5 148) and adapter are connected

between RF OUT and INPUT an d Align Now, TG has been run.

Specifications Supplemental Informati on
Warm- up 5 minutes
Specifications Supplemental Information
Output Frequency Range 9 kHz to 3.0 GHz
Specifications Supplemental Informati on
Minimum Resolution BW 1 kHz Not usable with resolution
bandwidths 300 Hz
(Option 1DR)
Specifications Supplemental Information
Output Power Level
Range 2 to 66 dBm
Resolution 0.1 dB
Absolute Accuracy (at 50 MHz
with coupled source
attenuator, referenced to
20 dBm)
± 0.75 dB
Verni er
Range 8 dB
Accuracy (with coupled
source attenuator, 50 MHz,
20 dBm)
Incremental ±0.2 dB/dB
Cumulative ±0.5 dB, total
Output Attenuator Range 0 to 56 dB in 8 dB steps