Chapter 1 15
Agilent E4403B Specifications and Characteristics
Amplitude
Agilent E4403B Specifications and
Characteristics
Amplitude
Amplitude specifications do not apply for the negative peak detector
mode.
Specifications Supplemental Information
Measurement Range Displayed Average Noise
Level to Maximum Safe Input
Level
Input Attenuator Range 0 to 65 dB, in 5 dB steps 0 to 75 dB, in 5 dB steps,
characteristic
Specifications Supplemental Information
Maximum Safe Input Level
Average Continuous Power +30 dBm (1 W)
(Input attenuator setting
5 dB)
Peak Pulse Power
(for <10 µsec pulse width,
<1% duty cycle, and input
attenuation 30 dB)
+50 dBm (100 W)
dc 100 Vdc
Specifications Supplemental Information
1 dB Gain Compression
Total power at input mixer a,b
50 MHz to 3.0 GHz 0 dBm
a. Mixer powe r leve l ( dB m) = input p ow er (dBm ) inp u t a ttenu ation (dB).
b. For resolution bandwidth s 1 kHz to 30 kHz, the maxim um in put signal amplitude must
be reference level +10 dB.